360" COILED MRI EXTENSION SET

Received Mar 5, 2019 · Event occurred Nov 3, 2018

Report 8392216 · MDR key 8392216

Device

Generic name

Tubing, Fluid Delivery

Product problems

  • Fluid/Blood Leak
  • Manufacturing, Packaging or Shipping Problem
  • Material Split, Cut or Torn

Patient

34675 DA

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

PER THE INITIAL REPORTER: THE MRI TUBING IS MANUFACTURED AND DELIVERED AS A LOOSE COIL ¿ SLINKY-LIKE. NORMALLY, DURING PREP, THE NURSES WILL PULL ON THE ENDS OF THE TUBING TO ALLOW THE TUBING TO STRETCH. THIS PARTICULAR LOT OF TUBING HAD THE COILS BONDED TOGETHER TOO STRONGLY. WHEN THE NURSE PULLED ON THE TUBING, IT RIPPED. ADDITIONALLY, THESE RIPS IN THE TUBING WERE HARD TO SPOT DURING THE PREP. THEY WEREN¿T NOTICED UNTIL THE TUBING LEAKED. THE SET WAS GIVEN TO CENTRAL SUPPLY FOR RETURN TO THE MANUFACTURER REP. UPDATE FROM THE MANUFACTURER PENDING.

Description of Event or Problem

PER THE INITIAL REPORTER: THE MRI TUBING IS MANUFACTURED AND DELIVERED AS A LOOSE COIL ¿ SLINKY-LIKE. NORMALLY, DURING PREP, THE NURSES WILL PULL ON THE ENDS OF THE TUBING TO ALLOW THE TUBING TO STRETCH. THIS PARTICULAR LOT OF TUBING HAD THE COILS BONDED TOGETHER TOO STRONGLY. WHEN THE NURSE PULLED ON THE TUBING, IT RIPPED. ADDITIONALLY, THESE RIPS IN THE TUBING WERE HARD TO SPOT DURING THE PREP. THEY WEREN¿T NOTICED UNTIL THE TUBING LEAKED.THE SET WAS GIVEN TO CENTRAL SUPPLY FOR RETURN TO THE MANUFACTURER REP. UPDATE FROM THE MANUFACTURER PENDING.