inforMED
MalfunctionMAL

HEMASHIELD GOLD KNITTED MICROVEL DOUBLE VELOUR

Received Mar 1, 2019 · Event occurred Jan 30, 2019

Report 1640201-2019-00013 · MDR key 8382219

Device

Generic name

Vascular Polyester Graft

Manufacturer

Intervascular Sas

Model number

M002020954080

Catalog number

M002020954080

Lot number

18F13

Product problems

  • Flaked
  • Flaked

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

DEVICE IS NOT ACCESSIBLE FOR TESTING AS IT REMAINED IMPLANTED IN PATIENT. A REVIEW OF THE COMPLAINT DEVICE HISTORY RECORDS, INDICATED THAT THE GRAFT WAS PROCESSED AND INSPECTED ACCORDING TO ESTABLISHED PROCEDURES AND WAS THEREFORE RELEASED FOLLOWING ACCEPTABLE QUALITY INSPECTIONS AND TESTS. THESE TESTS INCLUDE A 100% VISUAL INSPECTION OF THE GRAFT. SPECIFICALLY, THE REVIEW OF THE COLLAGEN COATING RECORDS EVIDENCED NO ANOMALY. A RETENTION SAMPLE FROM SAME LOT AND COATED ON THE SAME DAY AND UNDER THE SAME CONDITIONS AS THE INVOLVED DEVICE WAS IDENTIFIED. A VISUAL INSPECTION BY THE QA DEPARTMENT WILL BE PERFORMED. THE REVIEW OF POST-MARKETING HISTORICAL DATA INDICATED THAT NO OTHER SIMILAR COMPLAINT WAS REPORTED FOR THE SAME LOT NUMBER. THE INVESTIGATION IS STILL ONGOING. A FOLLOW-UP REPORT WILL BE SENT UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem

TWO VASCULAR GRAFTS WERE SUTURED TOGETHER AND WERE USED ON THE SAME PATIENT (SEE MANUFACTURER REPORT 1640201-2019-00012 FOR THE OTHER INVOLVED DEVICE). WHEN SUTURING THE GRAFTS TO BE IMPLANTED, IT WAS NOTICED THAT THE COLLAGEN COATING WAS CRUMBLING LIKE BREAD CRUMBS. THE SURGEON WET THE GRAFTS WITH STERILE SALINE AND NO MORE CRUMBLING WAS NOTICED. THE GRAFTS WERE IMPLANTED IN THE ABDOMINAL AREA WITH NO OTHER SIGNS OF ISSUES OR COMPLICATIONS.

Additional Manufacturer Narrative

(11/213) ONE RETENTION SAMPLE FROM SAME LOT AND COATED ON THE SAME DAY AND UNDER THE SAME CONDITIONS AS THE INVOLVED DEVICE WAS VISUALLY INSPECTED BY OUR QA SUPERVISOR. NO ANOMALY WAS FOUND. (4315) NO CONCLUSION CAN BE DRAWN SINCE BOTH INVOLVED PRODUCTS REMAINED IMPLANTED. HOWEVER, THE CONDUCTED INVESTIGATION SUGGESTS THAT THE DEVICES WERE NOT DEFECTIVE AT THE TIME OF MANUFACTURING.