XENON LIGHT SOURCE/FLASH

Received Feb 26, 2019 · Event occurred Oct 3, 2018

Report 8369613 · MDR key 8369613

Device

Generic name

Light Source, Photographic, Fiberoptic

Model number

20133120

Product problems

  • Material Discolored
  • Device Emits Odor
  • Failure to Power Up
  • Smoking
  • Material Deformation
  • Therapeutic or Diagnostic Output Failure

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

COMPLAINT OF DEVICE EMITTED SMOKE CONFIRMED BASED ON VISUAL OBSERVATIONS + ODOR PRESENT. DEVICE TURNED ON BUT WOULD NOT ILLUMINATE BULB. FOUND INTERNAL FUSE F1 OPEN ON XENON LAMP POWER SUPPLY BOARD. OBSERVATION OF COMPONENT CAPACITOR C4 ON XENON LAMP POWER SUPPLY INDICATES A FAILURE/THERMAL EVENT TOOK PLACE. COMPONENT IS DISCOLORED AND PLASTIC IS DISFIGURED. DEVICE WAS IN USE AT TIME OF EVENT IN OPERATING ROOM. A CODE RED ¿ FIRE/SMOKE RESPONSE BY STAFF AND EMERGENCY SERVICES TOOK PLACE.