inforMED
MalfunctionKSS

CENTRIFUGE 115V 50/60HZ DRUCKR

Received Feb 21, 2019 · Event occurred Feb 14, 2019

Report 0001825034-2019-00763 · MDR key 8357302

Device

Generic name

Blood Work Supplies

Catalog number

755VES

Lot number

60120511

Product problems

  • Device Difficult to Setup or Prepare
  • Device Difficult to Setup or Prepare

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Additional Manufacturer Narrative

(B)(4). THE DEVICE HAS BEEN RETURNED AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED THE CENTRIFUGE MALFUNCTIONED CAUSING THE BALANCING TO BE OFF, NOT ALLOWING FOR AN ACCURATE SPIN. THIS WAS IDENTIFIED IN AN OFFICE SETTING; NO PATIENT OR SURGICAL INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORTING.

Description of Event or Problem

NO FURTHER INFORMATION AVAILABLE AT THE TIME OF THIS REPORTING.

Additional Manufacturer Narrative

REPORTED EVENT WAS CONFIRMED VIA VISUAL EXAMINATION OF THE RETURNED DEVICE. THE DEVICE EXHIBITS DAMAGE CONSISTENT WITH USE OVER TIME. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. ROOT CAUSE IS ATTRIBUTED TO WEAR OUT DUE TO NORMAL USE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.