RF GROUNDING PAD DISPOSABLE WITH CORD
Received Feb 15, 2019 · Event occurred Jan 24, 2019
Report 3011270181-2019-00006 · MDR key 8345777
Device
Generic name
Standard Rf Cables & Accessories
Manufacturer
Avanos Medical Inc.Model number
PMA-GP-BAYCatalog number
109860601
Lot number
UNKNOWN
Product problems
- Excessive Heating
- Excessive Heating
Patient
Not reported
- Full thickness (Third Degree) Burn
- Full thickness (Third Degree) Burn
Narrative
Additional Manufacturer Narrative
THE ACTUAL COMPLAINT PRODUCT WAS NOT RETURNED FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD IS NOT POSSIBLE AS NO LOT NUMBER WAS PROVIDED. ALL INFORMATION REASONABLY KNOWN AS OF 15-FEB-2019 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL, INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL, INC. AVANOS MEDICAL, INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL, INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT COMP (B)(4).
Description of Event or Problem
IT WAS REPORTED THE PATIENT SUSTAINED A SKIN BURN UNDER THE GROUNDING PAD AREA OF THE SKIN. THE PATIENT WAS DIABETIC. THE PATIENT HAD A LEFT THREE LEVEL LUMBAR STANDARD RADIO FREQUENCY ABLATION (SRA) ON (B)(6) 2019. THE GROUNDING PAD WAS PLACED ON THE BACK OF THE UPPER LEFT THIGH. THE PATIENT WAS NOT HAIRY AND THE GROUNDING PAD ADHESION APPEARED NORMAL. THE PROCEDURE PROCEEDED WITHOUT INCIDENT WITH BOTH MOTOR AND SENSORY TESTING, NO HIGH IMPEDANCES NOR WARNINGS ON PAIN MANAGEMENT GENERATOR (PMG). SIX DAYS AFTER THE PROCEDURE THE PATIENT WENT TO THE EMERGENCY ROOM (ER) WITH REPORT OF A BURN AT GROUNDING PAD SITE. THE NURSE REVIEWED THE SKIN AND DESCRIBED THE SKIN AS A FULL THICKNESS CIRCULAR WOUND THE SIZE OF A DIME. IT IS UNKNOWN WHAT TREATMENT WAS DONE IN THE ER. NO ADDITIONAL INFORMATION WAS PROVIDED.