TRITON DTS PKG US STD

Received Jan 14, 2019 · Event occurred Dec 21, 2018

Report 9616086-2019-00004 · MDR key 8245034

Device

Generic name

Equipment, Traction, Powered

Manufacturer

Djo, Llc

Model number

2841

Product problems

  • Mechanical Problem
  • Mechanical Problem

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE TRACTION MACHINE "WILL PULL OR RELEASE ROPE WITHOUT ANY WARNING". THE EVENT REPORTEDLY OCCURRED WHILE TESTING THE MACHINE; THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

THE DEVICE IS NOT AVAILABLE FOR EVALUATION. IF THE DEVICE BECOMES AVAILABLE AS SUCH, A FOLLOW-UP REPORT WILL BE SUBMITTED ACCORDINGLY.

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED FOR EVALUATION. WE COULD NOT REPLICATE THE ROPE PULLING TOO HARD OR FEEDING OUT ON ITS OWN; THE SAFETY RISK COULD NOT BE CONFIRMED. THE DEVICE PASSED ALL FUNCTIONAL TESTING BUT WOULD NOT RETURN ALL THE WAY TO ZERO POUNDS; IT WENT DOWN TO 4 POUNDS. THIS WOULD NOT HAVE CONTRIBUTED TO THE INCIDENT OR HAVE BEEN A SAFETY RISK AS IT STILL RETURNED TO THE RANGE WHERE THE ROPE COULD RELEASE. THE MOTOR CONTROL CIRCUIT BOARD WAS REPLACED AND THE DEVICE WAS CALIBRATED. THE SPRING WAS REPLACED BECAUSE THE ROPE WOULD NOT RETRACT ALL THE WAY INTO THE DEVICE. THIS WOULD NOT HAVE CONTRIBUTED TO THE EVENT.