PORTEX CUFF INFLATOR PRESSURE GAUGE WITH CONNECTING TUBES

Received Jan 11, 2019

Report 3012307300-2019-00187 · MDR key 8241484

Device

Generic name

Monitor, Airway Pressure (includes Gauge And/or Alarm)

Catalog number

100/568/000

Lot number

3423245

Product problems

  • Output Problem
  • Output Problem

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

INFORMATION WAS RECEIVED THAT A SMITHS MEDICAL CUFF INFLATOR HAS A PENDANT POINTER. THE TIP IS TOUCHING THE CUFF INFLATOR SURFACE.

Additional Manufacturer Narrative

ONE CUFF INFLATOR PRESSURE GAUGE WAS RETURNED FOR EVALUATION. UPON VISUAL INSPECTION, DEVICE FOUND TO BE GOOD PHYSICAL CONDITION. DEVICE UNDERWENT FUNCTIONAL TESTING WHICH INCLUDED CLOSING THE INFLATION PORT WITH FINGER, WHILE PRESSING CUFF INFLATOR 30 TIMES REPEATEDLY. AS A RESULT, THE REPORTED CUSTOMER COMPLAINT WAS NOT CONFIRMED. WHILE NO DEFINITIVE ROOT CAUSE TO THE REPORTED ISSUE COULD BE DETERMINED, THIS INVESTIGATION REVEALED NO INTRINSIC EVIDENCE TO SUGGEST A ROOT CAUSE RELATED TO MANUFACTURING.