SYNGO DYNAMICS
Received Feb 2, 2007 · Event occurred Jan 17, 2007
Report 1836549-2007-00001 · MDR key 823438
Device
Generic name
Picture Archiving And Communication Sys
Manufacturer
Siemens Medical Solutions Usa, Inc.Model number
REV. 5.0Catalog number
NA
Lot number
NA
Product problems
- Incorrect Measurement
Patient
*
- Unknown (for use when the patient's condition is not known)
Narrative
Description of Event or Problem
A CUSTOMER SITE REPORTED THAT THE V MAX VALUE WAS INCORRECT WHEN USING THE SYSTEM'S CARDIAC CALCULATION FEATURE FOR SPECIFIC VTI MEASUREMENTS (AORTIC VALVE VELOCITY-TIME INTEGRAL). THE SYSTEM'S DERIVED V MAX WAS LOWER THAN EXPECTED. CONSEQUENTLY THE MEASUREMENTS IMPORTED INTO WORKSHEETS AND REPORTS WERE INCORRECT. THE FACTORY WAS ABLE TO REPRODUCE THIS REPORTED ISSUE. ALTHOUGH WE ARE UNAWARE OF ANY PT INJURY ASSOCIATED WITH THIS ISSUE, THIS PROBLEM HAS THE POTENTIAL TO AFFECT A PHYSICIAN'S DIAGNOSTIC DECISIONS IF IT RECURS. A CUSTOMER SAFETY ADVISORY NOTICE WAS CREATED AND WILL BE DISTRIBUTED TO ALL AFFECTED CUSTOMERS. FURTHERON A SOFTWARE PATCH WILL BE MADE AVAILABLE FOR UPGRADE OF THE SOFTWARE VERSION.
Remedial action
- Notification