inforMED
MalfunctionLMB

SYNGO DYNAMICS

Received Feb 2, 2007 · Event occurred Jan 17, 2007

Report 1836549-2007-00001 · MDR key 823438

Device

Generic name

Picture Archiving And Communication Sys

Model number

REV. 5.0

Catalog number

NA

Lot number

NA

Product problems

  • Incorrect Measurement

Patient

*

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

A CUSTOMER SITE REPORTED THAT THE V MAX VALUE WAS INCORRECT WHEN USING THE SYSTEM'S CARDIAC CALCULATION FEATURE FOR SPECIFIC VTI MEASUREMENTS (AORTIC VALVE VELOCITY-TIME INTEGRAL). THE SYSTEM'S DERIVED V MAX WAS LOWER THAN EXPECTED. CONSEQUENTLY THE MEASUREMENTS IMPORTED INTO WORKSHEETS AND REPORTS WERE INCORRECT. THE FACTORY WAS ABLE TO REPRODUCE THIS REPORTED ISSUE. ALTHOUGH WE ARE UNAWARE OF ANY PT INJURY ASSOCIATED WITH THIS ISSUE, THIS PROBLEM HAS THE POTENTIAL TO AFFECT A PHYSICIAN'S DIAGNOSTIC DECISIONS IF IT RECURS. A CUSTOMER SAFETY ADVISORY NOTICE WAS CREATED AND WILL BE DISTRIBUTED TO ALL AFFECTED CUSTOMERS. FURTHERON A SOFTWARE PATCH WILL BE MADE AVAILABLE FOR UPGRADE OF THE SOFTWARE VERSION.

Remedial action

  • Notification