ENDO GRASP
Received Dec 27, 2018 · Event occurred Nov 28, 2018
Report 2647580-2018-06320 · MDR key 8199927
Device
Generic name
Motor, Surgical Instrument, Pneumatic Powered
Manufacturer
Us Surgical Puerto RicoModel number
174233Catalog number
174233
Lot number
P6C0026X
Product problems
- Material Split, Cut or Torn
- Material Split, Cut or Torn
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
EVALUATION SUMMARY: POST MARKET VIGILANCE (PMV) LED AN EVALUATION OF ONE DEVICE. SHEARING DAMAGE OF THE PLASTIC CABLE SHEATH WAS OBSERVED. RECORDS FROM EACH MANUFACTURING LOT ARE THOROUGHLY REVIEWED TO ENSURE THAT PRODUCTS ARE RELEASED MEETING ALL QUALITY RELEASE SPECIFICATIONS AT THE TIME OF MANUFACTURE. REPLICATION OF THE REPORTED CONDITION MAY OCCUR WHERE AN EXCESSIVE FORCE OR LEVERAGE IS APPLIED TO THE DEVICE WHILE RETRACTING THE CLEVIS ASSEMBLY AND THE CLEVIS BOOT IS SUBSEQUENTLY DAMAGED. THIS SUBSEQUENT DAMAGE MAY RESULT IN IMPROPER RETRACTION AND DEPLOYMENT OF THE JAWS DURING ARTICULATION AND MAY CAUSE SHEARING OF THE PLASTIC CABLE SHEATH. THE ROOT CAUSE OF THE OBSERVED DAMAGE WAS MISUSE OF THE PRODUCT WHICH CAUSED OR CONTRIBUTED TO THE REPORTED CONDITION. NO FURTHER ACTIONS HAVE BEEN DEEMED NECESSARY AT THIS TIME. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Additional Manufacturer Narrative
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem
ACCORDING TO THE REPORTER DURING A CASE OF EXCISION OF MEDIASTINAL LESION AND THORACOSCOPY, BOTH DEVICE HAD THEIR COATING PEELED-OFF AFTER JUST ONE USE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT HARM.