ARCHITECT LH
Received Dec 27, 2006 · Event occurred Nov 2, 2006
Report 1415939-2006-00080 · MDR key 819961
Device
Generic name
Cmia For The Quantitative Determination Of Human Leuteinizin
Manufacturer
Abbott LaboratoriesModel number
NACatalog number
6C25-26
Lot number
33198M200
Product problems
- Other (for use when an appropriate device code cannot be identified)
- Failure to Calibrate
Patient
NO INFO
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
THE CUSTOMER ATTEMPTED A CALIBRATION OF ARCHITECT LH REAGENT LOT 33198M200 AND A CALIBRATION FAILURE 1227 (CAL F TOO HIGH) OCCURRED. NO SUSPECT PT RESULTS WERE GENERATED.
Additional Manufacturer Narrative
THIS ISSUE WAS IDENTIFIED WHILE INVESTIGATING AN INCREASE IN CUSTOMER COMPLAINTS FOR UPWARD SHIFTS IN NON-ABBOTT CONTROL AND PATIENT SAMPLE RESULTS, AS WELL AS CALIBRATION ERRORS. SPECIFICALLY, ERROR CODE 1227 "ASSAY (LH) NUMBER (641) CALIBRATION FALURE, FIT CONCENTRATION TOO HIGH FOR CAL F" MAY BE GENERATED, WHICH WOULD RESULT IN FAILURE TO OBTAIN A VALID CALIBRATION CURVE. ABBOTT HAS NOT REC'D ANY REPORTS OF ADVERSE EVENTS RELATED TO THE AFFECTED LOTS OF ARCHITECT LH REAGENTS. THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
Remedial action
- Recall