MalfunctionCAN
GAS CYLINDER PRESSURE REGULATOR
Received Dec 20, 2018 · Event occurred Jun 22, 2018
Report MW5082394 · MDR key 8189600
Device
Generic name
Gas Cylinder Pressure Regulator
Manufacturer
Airgas UsaProduct problems
- Break
- Device Emits Odor
- High Test Results
- Gas/Air Leak
Patient
23 YR
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Description of Event or Problem
ON (B)(6) 2018 SERVICE TECHNICIAN DISCOVERED ODOR AT CARBON DIOXIDE GAS MANIFOLD COMING FROM ONE CYLINDER. CO2 OUTLETS WERE TESTED. TESTS INDICATED ELEVATED LEVELS OF HYDROCARBONS IN THE C02 LINE, NOT MEETING NFPA99 STANDARDS. C02 REGULATOR AND PARTS OF MANIFOLD WERE FOUND DAMAGED AND WERE REPLACED. C02 WAS USED DURING LAPAROSCOPIC SURGERY FOR INSUFFLATION OR DURING CARDIOVASCULAR SURGERY FOR OXYGEN DISPLACEMENT/ SURGICAL FIRE PREVENTION.