inforMED
MalfunctionCAN

GAS CYLINDER

Received Dec 20, 2018 · Event occurred Jul 8, 2018

Report MW5082383 · MDR key 8189474

Device

Generic name

Regulator, Pressure, Gas Cylinder

Manufacturer

Airgas Usa

Product problems

  • Break
  • Device Emits Odor
  • High Test Results
  • Gas/Air Leak

Patient

15 YR

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

ON (B)(6) 2018, SERVICE TECH DISCOVERED ODOR AT CARBON DIOXIDE GAS MANIFOLD COMING FROM ONE CYLINDER. CO2 OUTLETS WERE TESTED. TESTS INDICATED ELEVATED LEVELS OF HYDROCARBONS IN THE CO2 LINE, NOT MEETING NFPA99 STANDARDS. CO2 REGULATOR AND PARTS OF MANIFOLD WERE FOUND DAMAGED AND WERE REPLACED. CO2 WAS USED DURING LAPAROSCOPIC SURGERY FOR INSUFFLATION OR DURING CARDIOVASCULAR SURGERY FOR OXYGEN DISPLACEMENT / SURGICAL FIRE PREVENTION.