inforMED
MalfunctionFIL

FRESENIUS DIALYSIS DELIVERY SYSTEM

Received Feb 1, 2007 · Event occurred Jan 2, 2007

Report 2937457-2007-00002 · MDR key 817697

Device

Generic name

Hemodialysis Machine

Model number

2008K

Catalog number

NA

Lot number

NA

Product problems

  • Output above Specifications
  • Unknown (for use when the device problem is not known)

Patient

65 YR

  • Low Blood Pressure/ Hypotension
  • Therapy/non-surgical treatment, additional

Narrative

Description of Event or Problem

A DIALYSIS FACILITY REPORTED THAT A MACHINE REMOVED TOO MUCH FLUID FROM A PT DURING A HEMODIALYSIS TREATMENT. THE PT BECAME HYPOTENSIVE AND WAS GIVEN A TOTAL OF 700 ML OF SALINE. BASED ON THE REPORTED PRE AND POST WEIGHTS, SALINE GIVEN AND WHAT THE MACHINE HAD INDICATED WAS REMOVED, THERE WAS A FLUID WEIGHT LOSS VARIANCE OF 1.9 KG. THERE WAS NO SERIOUS INJURY.