MalfunctionIPF

INTELECT MOBILE 2CH STIM INT'L STD

Received Dec 6, 2018 · Event occurred Nov 19, 2018

Report 9616086-2018-00025 · MDR key 8139663

Device

Generic name

Stimulator, Muscle, Powered

Manufacturer

Djo, Llc

Model number

2777

Product problems

  • Temperature Problem
  • Temperature Problem

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF EVALUATION.

Description of Event or Problem

IT WAS REPORTED THAT THE "DEVICE BURNED A HOLE ON THE SIDE OF THE UNIT." FURTHER INFORMATION HAS BEEN REQUESTED.

Additional Manufacturer Narrative

ONE INTELECT MOBILE 2-CHANNEL STIM INTERNATIONAL STANDARD UNIT WAS RETURNED FOR EVALUATION. THE POWER SUPPLY UNIT (PSU) WAS NOTED TO HAVE BEEN HELD BY ONE SCREW INSTEAD OF TWO AS NORMAL. THIS IS INDICATIVE OF INSUFFICIENT ELECTRICAL INSTALLATION BY THE END USER SITE. THE PSU WAS SHIFTED ON ONE SIDE. THE BURN STARTED BY THE 230 VOLT INPUT. AN INTERNAL METAL STRUCTURAL COMPONENT WAS FOUND TO HAVE RUSTED. THERE IS WHITE RESIDUE IN CONJUNCTION WITH THE RUST, WHICH INDICATES WATER INGRESS HAD OCCURRED. IT IS MOST LIKELY THAT WATER ENTERED THE UNIT, CAUSING A SHORT-CIRCUIT.