PRISMA CONTROL UNIT

Received Jan 12, 2007 · Event occurred Dec 14, 2006

Report 9616240-2007-00008 · MDR key 810527

Device

Generic name

Intensive Care Hemodialysis

Manufacturer

Gambro Dasco

Model number

PRISMA

Catalog number

018080001D

Lot number

NA

Product problems

  • Leak/Splash

Patient

YR

  • Blood Loss

Narrative

Additional Manufacturer Narrative

NO MEDICAL INTERVENTION WAS REPORTED. QUANTITY OF BLOOD LOSS UNK. THE MACHINE WAS CHECKED BY THE SERVICE TECH. HE FOUND THERE WAS A 3-WAY STOP-COCK CONNECTOR BETWEEN THE BLOOD TUBING SET AND THE CATHETER AS A POTENTIAL SOURCE OF THE BLOOD LEAKAGE. A GAMBRO SERVICE TECH INSPECTED THE MACHINE. NO MALFUNCTION WAS FOUND. THE TECHNICIAN WAS ABLE TO SIMULATE AND DEMONSTRATE THE PROPER FUNCTION OF THE PRESSURE TRENDING LIMITS: "CAUTION: RETURN DISCONNECTION CAN NOT BE DETECTED" ALARM AND 'WARNING: RETURN DISCONNECTION ALARM". THE SECOND ALARM STOPS THE BLOOD PUMP AND CLOSES THE VENOUS CLAMP WHEN THE RETURN PRESSURE FALLS BELOW + 10 MMHG. THE PHYSICIAN COMPLAINED THAT DURING THE "CAUTION: RETURN DISCONNECTION CAN NOT BE DETECTED" ALARM A SIGNIFICANT AMOUNT OF BLOOD COULD BE LOST.

Description of Event or Problem

THE CUSTOMER REPORTED THERE WAS A LEAKAGE ON THE RETURN LINE WITH A BLOOD LOSS WITH NO ALARMS.