OSCILLATING ULTRASONIC PROBE
Received Nov 21, 2018 · Event occurred Oct 23, 2018
Report 9610617-2018-00091 · MDR key 8096620
Device
Generic name
Oscillating Ultrasonic Probe
Manufacturer
Karl Storz Se & Co. KgModel number
27093LL-SCatalog number
27093LL-S
Lot number
ST01
Product problems
- Material Separation
- Material Separation
Patient
62 YR
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
THE DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION; HOWEVER, OUR SALES REPRESENTATIVE VIEWED THE DEVICE AND ALSO PROVIDED A PICTURE, WHICH SHOWS THREE PIECES WERE BROKEN OFF OF THE OSCILLATING PROBE TIP. THE HOSPITAL CONFIRMED THAT THE BROKEN PIECES WERE RETRIEVED AND THERE WAS NO HARM TO PATIENT. IN ADDITION, WE LEARNED FROM THE HOSPITAL THAT THE KIDNEY STONE WAS VERY HARD AND THAT AFTER THE SURGEON HAD BEEN WORKING ON IT FOR SOME TIME USING THIS PROBE, THE ULTRASONIC CALCUSON UNIT DISPLAYED A WARNING LIGHT; HE CONTINUED USING THE PROBE AND THE TIP EVENTUALLY BROKE. IT IS POSSIBLE THE WARNING LIGHT WAS THE TRANSDUCER FAULT INDICATOR - RESONANCE WARNING, WHICH CAN INDICATE EITHER A DAMAGED PROBE OR TRANSDUCER. A POSSIBLE ROOT CAUSE OF THIS TYPE OF BREAKAGE IS OVERHEATING AND/OR MECHANICAL OVERLOAD. BOTH ARE USER RELATED ISSUES AND ARE ADDRESSED IN OUR IFU.
Description of Event or Problem
ALLEGEDLY, DURING A PERCUTANEOUS NEPHROLITOMY PROCEDURE, THREE PIECES OF TIP OF THE ULTRASONIC PROBE BROKE OFF INTO THE PATIENT. THE BROKEN PIECES WERE RETRIEVED AND PROCEDURE WAS COMPLETED USING A DIFFERENT PROBE. HOSPITAL CONFIRMED THERE WAS NO INJURY TO PATIENT.