Received Nov 16, 2018 · Event occurred Oct 25, 2018
Report 3032391-2018-00018 · MDR key 8079169
Device
Generic name
Carpal Tunnel Release Blade Assembly, (white)
Manufacturer
Medline RenewalCatalog number
81010R
Lot number
375194
Product problems
- Material Fragmentation
Patient
Not reported
- Foreign Body In Patient
- Foreign Body In Patient
Narrative
Additional Manufacturer Narrative
IT WAS REPORTED THAT THROUGH EXAMINATION UNDER A MICROSCOPE, MICROSCOPIC BLACK SPECKS WERE VISUALIZED IN THE POST-CARPAL TUNNEL SURGERY WOUND. IT WAS SUSPECTED THAT THE MICROSCOPIC BLACK SPECKS WERE FROM THE REPROCESSED MICROAIRE'S CARPAL TUNNEL RELEASE BLADE ASSEMBLY. PER REPORT, THESE MICROSCOPIC BLACK SPECKS WERE SUCCESSFULLY REMOVED FROM THE PATIENT THROUGH AN UNIDENTIFIED METHOD. THERE WAS NO REPORT OF PROLONGED ANESTHESIA, A SERIOUS INJURY OR ANOTHER ADVERSE PATIENT EVENT RELATED TO THE REPORTED INCIDENT. DUE TO THE REPORTED INCIDENT AND IN AN ABUNDANCE OF CAUTION, THIS MEDWATCH IS BEING FILED. THE SAMPLE WAS NOT AVAILABLE TO BE RETURNED FOR EVALUATION. COMPANION SAMPLES FROM THE REPORTED LOT NUMBER WERE RETURNED AND THE REPORTED COMPLAINT IS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED AT THIS TIME. NO ADDITIONAL INFORMATION IS AVAILABLE. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
Description of Event or Problem
IT WAS REPORTED THAT MICROSCOPIC BLACK SPECKS FROM THE REPROCESSED MICROAIRE'S CARPAL TUNNEL RELEASE BLADE ASSEMBLY WERE NOTED IN THE WOUND POST CARPAL TUNNEL SURGERY.