inforMED
MalfunctionJIL

CHEMSTRIP ® 10 MD

Received Nov 12, 2018 · Event occurred Oct 22, 2018

Report 1823260-2018-04180 · MDR key 8059677

Device

Generic name

Urine Test Strips

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

03260763160

Lot number

27638305

Product problems

  • Low Test Results
  • Low Test Results

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

THE CUSTOMER COMPLAINED OF FALSE NEGATIVE NITRITE AND ERYTHROCYTES RESULTS FOR PATIENT URINE SAMPLES FROM URISYS 1100 URINE ANALYZER SERIAL NUMBER (B)(4). THE NITRITE RESULT FROM THE METER WAS NEGATIVE AND THE RESULT WAS POSITIVE WITH THE VISUAL TEST. THE ERYTHROCYTES RESULT FROM THE METER WAS NEGATIVE AND THE RESULT WAS POSITIVE WITH THE VISUAL TEST. THE SPECIFIC DATE OF THE MEASUREMENTS WAS NOT PROVIDED. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE QC WAS ACCEPTABLE. THE SUSPECT PRODUCT WAS REQUESTED TO BE RETURNED FOR INVESTIGATION. THE RETENTION MATERIAL OF LOT 27638300 WAS MEASURED ON A URISYS 1800 ANALYZER AND WAS VISUALLY CHECKED WITH NATIVE URINE AND A NITRITE DILUTION SERIES. THE RESULTS OF THE MEASUREMENTS FULFILL OUR REQUIREMENTS. NO FALSE NEGATIVE RESULTS WERE OBSERVED.

Additional Manufacturer Narrative

THE RETENTION MATERIAL OF LOT 27638300 WAS MEASURED ON A COBAS U411 AND WAS VISUALLY CHECKED WITH NATIVE URINE AND A NITRITE DILUTION SERIES. THE RESULTS OF THE MEASUREMENTS FULFILL OUR REQUIREMENTS. NO FALSE NEGATIVE RESULTS WERE OBSERVED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Additional Manufacturer Narrative

THE CUSTOMER'S METER AND TEST STRIPS WERE PROVIDED FOR INVESTIGATION. THE METER WAS RETURNED WITHOUT A TEST STRIP TRAY. THE METER WAS CLEAN AND NO DAMAGE. THE RETURNED TEST STRIPS HAD NO DISCOLORATIONS OR ABNORMALITIES. DURING INTERNAL TESTING, ROCHE DETERMINED THE LIMITS OF DETECTION (LOD) FOR PROTEIN, NITRITE, LEUKOCYTES, AND ERYTHROCYTES ON THE URISYS 1100 URINE ANALYZER WITH CHEMSTRIP® 5 OB, CHEMSTRIP 7, CHEMSTRIP 10 MD, AND CHEMSTRIP 10 UA TEST STRIPS WERE HIGHER THAN WHAT IS LISTED IN THEIR RESPECTIVE TEST STRIP METHOD SHEETS. THIS CAN LEAD TO FALSE NEGATIVE RESULTS ON THE URYSIS 1100 URINE ANALYZER FOR THE FOUR AFFECTED PARAMETERS. ROCHE HAS PROVIDED CUSTOMERS WITH THE FOLLOWING INSTRUCTIONS AND WORKAROUNDS: CHEMSTRIP 5 OB AND CHEMSTRIP 7 TEST STRIPS SHOULD ONLY BE READ VISUALLY. THEY CAN NO LONGER BE READ ON THE URISYS 1100 URINE ANALYZER. CHEMSTRIP 10 MD OR CHEMSTRIP 10 UA TEST STRIPS CAN BE USED WITH THE URISYS 1100 URINE ANALYZER; HOWEVER, A NEGATIVE RESULT FOR ANY ONE OF THE FOUR AFFECTED PARAMETERS (I.E., PROTEIN, NITRITE, LEUKOCYTES, AND ERYTHROCYTES) ON THE URISYS 1100 URINE ANALYZER MUST BE REPEATED WITH A NEW TEST STRIP THAT IS READ VISUALLY.