inforMED
MalfunctionFSH

JAMSHIDI NEEDLE BONE MARROW

Received Oct 30, 2018 · Event occurred Oct 9, 2018

Report 9680904-2018-00026 · MDR key 8019340

Device

Generic name

Biopsy Needles & Trays

Manufacturer

Carefusion, Inc

Catalog number

DJ4011X

Lot number

0001220841

Product problems

  • Unsealed Device Packaging

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

(B)(4) UPON COMPLETION OF CAREFUSIONS INVESTIGATION A FOLLOW UP SUBMISSION WILL BE COMPLETED.

Description of Event or Problem

UPON INSPECTION OF OUR QC PERSONNEL AND PLACEMENT OF REGULATORY LABEL TO EACH ITEMS, 5 CASES OF (B)(4) CONTAIN ITEMS WITH DEFECTIVE SEAL.

Additional Manufacturer Narrative

(B)(4) OUR QUALITY ENGINEERS RECEIVED EIGHT PHOTOS FOR EVALUATION. DURING VISUAL EVALUATION, IT WAS DETERMINED ALL NINETEEN (19) POUCHES WERE OPENED ON THE SIDES, THEREFORE, FAILURE MODE COULD BE CONFIRMED. ALTHOUGH REPORTED FAILURE WAS CONFIRMED THROUGH VISUAL ANALYSIS A ROOT CAUSE COULD NOT BE DETERMINED THROUGH THIS INVESTIGATION. TO FURTHER INVESTIGATE ROOT CAUSE, BD HAS INITIATED A CAPA. THE FAILURE MODE WILL BE ENTERED INTO THE COMPLAINT MANAGEMENT SYSTEM TO BE TRACKED/TRENDED. CUSTOMER COMPLAINT TRENDS ARE EVALUATED ON A MONTHLY BASIS DURING OUR QUALITY TEAM MEETING.

Description of Event or Problem

UPON INSPECTION OF OUR QC PERSONNEL AND PLACEMENT OF REGULATORY LABEL TO EACH ITEMS, 5 CASES OF DJ4011X CONTAIN ITEMS WITH DEFECTIVE SEAL.