inforMED
InjuryMAX

HONOUR SPACER SYSTEM

Received Oct 24, 2018 · Event occurred Oct 3, 2017

Report 3007922509-2018-00005 · MDR key 8002286

Device

Generic name

Honour Spacer System

Manufacturer

Nexxt Spine Llc

Model number

21-229-10

Catalog number

21-229-10

Product problems

  • Migration

Patient

50 YR

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

PEEK IMPLANT OF HEIGHT 10MM MIGRATED AS NOTICED IN AN X-RAY IN 2 WEEK POST-OP FOLLOW UP VISIT. SURGEON PERFORMED A REVISION SURGERY TO REPLACE THE 10MM IMPLANT WITH AN 11MM IMPLANT AS A BETTER FIT.

Remedial action

  • Other