ARCHITECT LH

Received Dec 13, 2006

Report 1415939-2006-00068 · MDR key 796963

Device

Generic name

Cmia For The Quantitative Determination Of Human Luteinizing

Model number

NA

Catalog number

6C25-22

Lot number

33198M200

Product problems

  • Use of Incorrect Control/Treatment Settings
  • Other (for use when an appropriate device code cannot be identified)

Patient

NO INFO

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

THE CUSTOMER OBSERVED AN UPWARD SHIFT IN ARCHITECT I2000 LH CONTROL VALUES USING LH REAGENT LOT 33198M200. THERE IS NO REPORT OF SUSPECT PATIENT RESULTS.

Additional Manufacturer Narrative

THIS ISSUE WAS IDENTIFIED WHILE INVESTIGATING AN INCREASE IN CUSTOMER COMPLAINTS FOR UPWARD SHIFTS IN NON-ABBOTT CONTROL AND PATIENT SAMPLE RESULTS, AS WELL AS CALIBRATION ERRORS. SPECIFICALLY, ERROR CODE 1227 "ASSAY (LH) NUMBER (641) CALIBRATION FAILURE, FIT CONCENTRATION TOO HIGH FOR CAL F" MAY BE GENERATED, WHICH WOULD RESULT IN FAILURE TO OBTAIN A VALID CALIBRATION CURVE. ABBOTT HAS NOT RECIEVED ANY REPORTS OF ADVERSE EVENTS RELATED TO THE AFFECTED LOTS OF ARCHITECT LH REAGENTS. AN INVESTIGATION IS IN PROCESS.

Remedial action

  • Recall