BARD PORT AND CATHETER IMPLANTED SUBCUTANEOUS
Received Oct 10, 2018 · Event occurred Sep 25, 2018
Report MW5080479 · MDR key 7957028
Device
Generic name
Mediport Ii Model Mp-dl Vascular Access Port
Manufacturer
Bard / C. R. Bard, Inc.Model number
2864Product problems
- Device Dislodged or Dislocated
Patient
48 YR
- Foreign Body In Patient
- Device Embedded In Tissue or Plaque
- Foreign Body In Patient
- Device Embedded In Tissue or Plaque
Narrative
Description of Event or Problem
A (B)(6) Y/O FEMALE PT WITH HISTORY OF LEFT BREAST CANCER, WHO IS S/P MEDIPORT INSERTION EARLIER THIS YEAR IN (B)(6). SHE HAS BEEN UNDERGOING NEOADJUVANT CHEMOTHERAPY. TODAY, IT WAS NOTED BY HER MEDICAL ONCOLOGIST THAT THE CATHETER DID NOT HAVE A GOOD BACKFLOW. ACCORDING TO DR (B)(6) (ONCOLOGIST), THE CATHETER WAS FLUSHED WITH SALINE, AND HE IMMEDIATELY NOTICED SLIGHT SWELLING IN THE SUBCUTANEOUS POCKET AREA. THE PT WAS THUS BROUGHT IN FOR WOUND EXPLORATION, FLUOROSCOPY, EVALUATION/REVISION OF THE MEDIPORT, POSSIBLE REMOVAL. DURING THE SURGERY, IT WAS FOUND THAT THE CATHETER WAS DISRUPTED BETWEEN 5 AND 6 CM FROM THE TIP, AND ON FLUOROSCOPIC VIEW, THE CATHETER TIP WAS FOUND TO BE RETAINED IN THE RIGHT ATRIUM. THE REMAINING CATHETER AND PORT WERE REMOVED AND TRANSFERRED TO HIGHER LEVEL OF CARE FOR FOREIGN BODY REMOVAL.