ELECSYS ANTI-CCP IMMUNOASSAY
Received Oct 11, 2018 · Event occurred Sep 19, 2018
Report 1823260-2018-03560 · MDR key 7955597
Device
Generic name
Antibodies, Anti-cyclic Citrullinated Peptide (ccp)
Manufacturer
Roche DiagnosticsModel number
NACatalog number
05031656190
Lot number
34774000
Product problems
- High Test Results
- High Test Results
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
CALIBRATION DATA AND QC DATA ARE WITHIN EXPECTATIONS. A GENERAL REAGENT ISSUE MOST LIKELY CAN BE EXCLUDED. THE DETECTION OF ANTI-CCP ANTIBODIES IS PERFORMED WITH A MIXTURE OF VARIOUS CCP'S. AS DIFFERENT METHODS USE DIFFERENT MIXTURES INCLUDING DIFFERENT (AMOUNTS) OF THESE PEPTIDES, IT IS VERY LIKELY THAT SPECIFICITY BETWEEN METHODS DIFFERS. THIS IS NOTED IN THE PRODUCT LABELING. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.
Additional Manufacturer Narrative
THIS EVENT OCCURRED IN (B)(6).
Description of Event or Problem
THE CUSTOMER COMPLAINED OF A DISCREPANT ELECSYS ANTI-CCP IMMUNOASSAY RESULT FOR 1 PATIENT SAMPLE ON COBAS 8000 E 602 MODULE SERIAL NUMBER (B)(4). THE PATIENT HAD GONE TO THE RHEUMATOLOGIST FOR PAIN. THE INITIAL ANTI-CCP RESULT WAS 213.3 U/L ON THE E 602 MODULE. THE PHYSICIAN QUESTIONED THE RESULT AND THE SAMPLE WAS SENT TO ANOTHER LABORATORY. THE ANTI-CCP RESULT WAS NEGATIVE WITH THE PHARMACIA IMMUNOCAP METHOD. THE ANTI-CCP RESULT WAS 0 U/L WITH THE EURO DIAGNOSTICA ELISA METHOD. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE RHEUMATOLOGIST STATED THAT PATIENTS' ANTI-CCP RESULTS ARE OFTEN POSITIVE WITH ROCHE TESTS AND NEGATIVE WITH OTHER METHODS.