BD AFFIRM¿ VPIII AMBIENT TEMPERATURE TRANSPORT SYSTEM
Received Oct 9, 2018 · Event occurred Sep 13, 2018
Report 9610847-2018-00327 · MDR key 7946720
Device
Generic name
Microbiological Specimen Collection And Transport Device
Manufacturer
Bd Infusion Therapy Systems Inc. S.a. De C.v.Catalog number
446255
Product problems
- Material Protrusion/Extrusion
Patient
Not reported
- Exposure to Body Fluids
- Laceration(s)
- Exposure to Body Fluids
- Laceration(s)
Narrative
Additional Manufacturer Narrative
MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. NO LOT # PROVIDED. DEVICE MANUFACTURE DATE: UNKNOWN. INVESTIGATION SUMMARY: BD MOLECULAR QUALITY INVESTIGATED AN END USER, AFFIRM ATTS GLASS USER INJURY (FINGER PUNCTURE) DURING DISPENSE OF SPECIMEN AT AN AFFIRM TESTING FACILITY. IT WAS CONFIRMED THAT THE END USER FOUND GLASS IN THE SPECIMEN COLLECTION TUBE. IT WAS DETERMINED THAT THE AFFIRM COLLECTION SITE WAS NOT FOLLOWING THE AFFIRM ATTS PACKAGE INSERT OR AFFIRM COLLECTION POSTER INSTRUCTIONS. THE COLLECTION SITE WAS CRUSHING THE ATTS GLASS AMPULE (AS INSTRUCTED), OPENING THE ATTS CAP AND POURING THE GLASS INTO THE AFFIRM SAMPLE COLLECTION TUBE (NOT INSTRUCTED). THE AFFIRM PACKAGE INSERT AND AFFIRM COLLECTION POSTER DOES NOT INSTRUCT COLLECTION SITES TO OPEN THE ATTS VIAL AND POUR THE GLASS INTO THE COLLECTION TUBE. BD HAS COMPLETED AN UPDATE TO THE AFFIRM COLLECTION POSTER TO PROVIDE FURTHER CLARIFICATION ON THE INSTRUCTIONS FOR USE WITH AFFIRM ATTS. BD MOLECULAR QUALITY WILL CONTINUE TO CLOSELY MONITOR FOR TRENDS ASSOCIATED WITH AFFIRM ATTS GLASS INJURY.
Description of Event or Problem
IT WAS REPORTED THAT WHILE USING A BD AFFIRM¿ VPIII AMBIENT TEMPERATURE TRANSPORT SYSTEM THE TECHNICIAN WAS PLACING THE SAMPLE TUBE INTO THE DEVICE WHEN SHARP PIECES OF THE GLASS AMPULE PENETRATED THE WALL OF THE DEVICE AND CUT THE FINGER OF THE TECHNICIAN RESULTING IN EXPOSURE TO PATIENT SAMPLE. THE TECHNICIAN RECEIVED A MEDICAL EVALUATION AND WAS TREATED FOR EXPOSURE TO BLOODBORNE PATHOGENS AT AN URGENT CARE FACILITY.