Received Dec 9, 2004

Report 8010047-2004-10054 · MDR key 793968

Device

Generic name

Sigmoidovideoscope

Model number

CF-Q160S

Catalog number

NA

Lot number

NA

Product problems

  • Image Display Error/Artifact

Patient

UNKNOWN

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

THE HOSPITAL REPORTED THEY EXPERIENCED A TOTAL LOSS OF IMAGE DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED WITH THE USE OF ANOTHER SIMILAR DEVICE. THERE WAS NO ALLEGATION OF HARM.

Additional Manufacturer Narrative

THE DEVICE IN QUESTION WAS RETURNED TO OLYMPUS FOR INVESTIGATION. THE INVESTIGATION REVEALED EVIDENCE OF FLUID INVASION IN THE ELECTRICAL CONNECTOR AND BODY CONTROL UNIT WHICH MOST LIKELY CAUSED THE USER'S EXPERIENCE. OLYMPUS ATTRIBUTED THESE PHENOMENONS TO IMPROPER REPROCESSING OF THE DEVICE.