MEDFUSION 2001I SYRINGE INFUSION PUMP
Received Dec 1, 2006 · Event occurred Oct 28, 2006
Report 2183502-2006-00175 · MDR key 793742
Device
Generic name
Syringe Infusion Pump
Manufacturer
Smiths Medical Md, Inc.Model number
2010ICatalog number
2010I
Lot number
NA
Product problems
- Excess Flow or Over-Infusion
- Repair
Patient
UNKNOWN
- Overmedicated
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
IT WAS REPORTED BY THE BIO-MED TECH AT THE USER FACILITY THAT THE DEVICE "OVERINFUSED." ADD'L INFO WAS REQUESTED AND REC'D FROM THE USER FACILITY'S RISK MANAGER WHO REPORTED THAT THERE WAS NO HARM TO THE PT AND THAT "THE MEDICATION WAS GIVEN AT A FASTER RATE THAN THE PUMP INDICATES SHOULD HAVE BEEN GIVEN." ADDITIONALLY THEY DID PERFORM A SHORT TEST WITH WATER IN THE SYRINGE AND IT APPEARED TO BE WORKING, BUT DID NOT PERFORM ANY LONG TERM TESTING.
Additional Manufacturer Narrative
THE SUSPECT DEVICE WAS RETURNED AND EVALUATED. AN ATTEMPT WAS MADE TO OPERATE THE PUMP TO TEST THE DELIVERY ACCURACY, THIS WAS NOT POSSIBLE DUE TO NUMEROUS ISSUES WITH THE DEVICE. THE TAMPER STICKER WAS NOT INTACT, THE DEVICE HAD BEEN SERVICED IN THE FIELD AND THERE WERE MULTPLE INDICATIONS THAT THE PUMP WAS NOT REASSSEMBLED CORRECTLY. ADDITIONALLY, THERE WAS PHYSICAL DAMAGE AND FLUID INGRESSION AND THE DEVICE WAS OUT OF CALIBRATION. THE DEVICE WAS EVALUATED AND THE ALLEGED FAILURE COULD NOT BE DUPLICATED AS THE DEVICE WAS REC'D IN AN INOPERABLE CONDITION.