CELL-DYN 22 CALIBRATOR
Received Dec 1, 2006
Report 2919069-2006-00106 · MDR key 789591
Device
Generic name
Cd22 Calibrator
Manufacturer
Abbott Diagnostics Div/celldynModel number
NACatalog number
99120-01
Lot number
3099 AND 3100
Product problems
- Nonstandard Device
- Low Test Results
Patient
NA
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
THE ACCOUNT STATED THAT THEY HAD CALIBRATED THEIR CD3200 SL ANALYZER WITH CELL-DYN 22 CALIBRATOR LOT #'S 3099 AND 3100. THE ACCOUNT HAS BEEN GETTING DELTA CHECKS ON SOME PTS. DUE TO THE RECALL LETTER THAT THEY RECEIVED THEY ARE CONCERNED. THE CUSTOMER TECHNICAL ADVOCATE (CTA) SUGGESTED THAT THE ACCOUNT CHECK THE PT PLATELET VALUES WITH AN OUTSIDE LABORATORY. THE ACCOUNT DECLINED. THE ACCOUNT STATED THAT THEY HAVE NOT HAD ANY ISSUES WITH QC.
Additional Manufacturer Narrative
ON-MARKET INSTABILITY FOR PLATELETS WITH CELL-DYN 22 CALIBRATOR LOTS 3098 AND 3099 IS BEING EXPERIENCED. THE CALIBRATOR COULD READK HIGHER BY AT LEAST 10% ON THE CELL-DYN RUBY AND AS MUCH AS 22% ON THE CELL-DYN 3200. IF THIS CALIBRATOR IS USED AND THE CALIBRATION FACTOR ON THE INSTRUMENT IS ADJUSTED, A NEGATIVE PLATELET SAMPLE BIAS COULD RESULT. AS A PRECAUTION, LOT 3100 WAS ALSO REMOVED FROM USE. INVESTIGATION INTO THE CAUSE OF THIS INSTABILITY IS IN PROGRESS. THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
Remedial action
- Recall