MalfunctionCGL

I-STAT CREA CARTRIDGE

Received Sep 6, 2018 · Event occurred Aug 8, 2018

Report 2245578-2018-00226 · MDR key 7850120

Device

Generic name

Crea Cartridge

Model number

NA

Catalog number

03P84-25

Lot number

A18139

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

75 YR

  • No Information
  • No Information

Narrative

Additional Manufacturer Narrative

(B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem

ON (B)(6) 2018, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CREA+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT CREATININE RESULT ON A (B)(6) MALE PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. (B)(6). THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

Additional Manufacturer Narrative

APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 09/17/2018. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE LOT PASSED FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AC, APPENDIX 1- PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED.