PORTEX® SELDINGER CHEST DRAINAGE KIT
Received Aug 29, 2018 · Event occurred Aug 2, 2018
Report 3012307300-2018-03568 · MDR key 7828393
Device
Generic name
Thoracic Catheter Insertion Tray
Manufacturer
Smiths Medical Asd; Inc.Catalog number
200/910/120
Lot number
3566212
Product problems
- Crack
- Fluid/Blood Leak
- Crack
- Fluid/Blood Leak
Patient
Not reported
- Fistula
- Fistula
Narrative
Description of Event or Problem
INFORMATION WAS RECEIVED INDICATING THAT AN AIR FISTULA WAS DETECTED LEADING TO LEAKING OF THE SMITHS MEDICAL PORTEX® SELDINGER CHEST DRAINAGE SYSTEM. THE DRAIN WAS REMOVED AS THE LEAK WAS UNABLE TO BE FOUND. UPON VISUAL EXAMINATION OF THE DRAIN, A 1 CM LONG LONGITUDINAL CRACK WAS NOTED AT THE SCREW ATTACHMENT OF THE DRAIN. IT WAS REPORTED THAT THE DRAINAGE SYSTEM WAS NOT REMOVING CEREBRAL SPINAL FLUID (CSF) AS IT WAS SUPPOSED TO BUT AFTER THE SYSTEM CHANGE OUT, 100MLS OF CSF DRAINED. THERE WERE NO REPORTED ADVERSE EFFECTS.
Additional Manufacturer Narrative
EVALUATION RESULTS: ONE USED PLEURAL CATHETER WAS RETURNED FOR INVESTIGATION. THE INVESTIGATORS CONFIRMED THE CUSTOMER REPORTED PROBLEM REGARDING THE 1 CM LONG LONGITUDINAL CRACK AT THE SCREW ATTACHMENT OF THE DRAIN. THE DEVICE HISTORY RECORD WAS REVIEWED AND SHOWED THAT THIS DEVICE MET ALL MANUFACTURING SPECIFICATION FOR PRODUCT RELEASED FOR DISTRIBUTION. NO ISSUES WERE IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT. THE INVESTIGATORS CONCLUDED THAT THE INCIDENT WAS NOT DUE TO A PRODUCT FAILURE. AN EXACT ROOT CAUSE FOR THE REPORTED FAILURE WAS NOT DETERMINED HOWEVER.