AMPLICOR CT/NQ TEST FOR C TRACHOMATIS
Received Nov 15, 2006
Report 2243471-2006-00013 · MDR key 782000
Device
Generic name
In Vitro Diagnostic Nucleic Acid Test
Manufacturer
Roche Molecular Systems, Inc.Model number
*Catalog number
20759392018
Lot number
NA
Product problems
- False Negative Result
Patient
*
- Unknown (for use when the patient's condition is not known)
Narrative
Description of Event or Problem
CUSTOMER REPORTED A SERIES OF SPECIMENS THAT WERE NEGATIVE FOR CHLAMYDIA TRACHOMATIC WITH THE ROCHE MOLECULAR SYSTEMS AMPLICOR CT/NG TEST, BUT POSITIVE WITH AN ALTERNATE METHOD.
Additional Manufacturer Narrative
THE CUSTOMER PROVIDED ROCHE MOLECULAR SYSTEMS WITH SAMPLES FROM THE SPECIMENS IN QUESTION. ROCHE MOLECULAR SYSTEMS, INC. INVESTIGATED THE CUSTOMER'S ALLEGATIONS AND CONFIRMED THAT THE SPECIMENS TESTED NEGATIVE FOR CHLAMYDIA TRACHOMATIS. FURTHER INVESTIGATION DETERMINED THAT THE SPECIMENS CONTAINED A MUTATED STRAIN OF C. TRACHOMATIS. THIS MUTATION RESULTED IN THE DELETION OF APPROX 350 BASE PAIRS FROM THE REGION WITHIN THE CRYPTIC PLASMID DNA COVERED BY THE TEST'S PRIMERS/AND OR PROBE. RMS IS CONTINUING OUR INVESTIGATION TO BETTER UNDERSTAND THE PREVALENCE AND GEOGRAPHIC DISTRIBUTION OF THE MUTATED STRAIN. THE PACKAGE INSERT WILL ALSO BE MODIFIED TO REFLECT THE RARE, BUT POSSIBLE OCCURRENCE OF MUTATIONS THAT COULD IMPACT THE TEST'S PERFORMANCE.
Remedial action
- Notification