INFANT HEATED WIRE CIRCUIT
Received Aug 8, 2018 · Event occurred Jun 23, 2018
Report 8030673-2018-00008 · MDR key 7762988
Device
Generic name
Oxygen Administration Kit
Manufacturer
Vyaire MedicalCatalog number
AH265
Lot number
0001119873
Product problems
- Moisture or Humidity Problem
- Moisture or Humidity Problem
Patient
Not reported
- Decreased Respiratory Rate
- Decreased Respiratory Rate
Narrative
Additional Manufacturer Narrative
AT THIS TIME, VYAIRE MEDICAL HAS NOT RECEIVED THE SAMPLE FOR INVESTIGATION. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem
THE RESPIRATORY THERAPIST REPORTED: RAIN-OUT IS A BIG PROBLEM ON INTUBATED BABY. DRAINING CIRCUIT VERY FREQUENTLY. RAIN -OUT IS SETTING OFF HIGH RESPIRATION RATE ALARM. CONNECTION AT THE SWIVEL FALLS OUT EASILY.
Additional Manufacturer Narrative
THE DEVICE SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION BY OUR QUALITY TEAM. HOWEVER, THE QUALITY TEAM IDENTIFIED THAT A DEFECTIVE CRIMPING COULD CAUSE BAD ELECTRICAL CONNECTION OF THE HEATING WIRE TO THE POWER SUPPLY. THIS COULD GENERATE RAIN OUT IN THE CIRCUIT. IF SET UP TO CRIMPING EQUIPMENT IS NOT PROPERLY DONE, THE CRIMPING HEIGHT CAN BE OUT OF SPEC.