AIRLIFE¿ INFANT HEATED WIRE CIRCUIT

Received Aug 8, 2018 · Event occurred Jun 24, 2018

Report 8030673-2018-00009 · MDR key 7762847

Device

Generic name

Oxygen Administration Kit

Manufacturer

Vyaire Medical

Catalog number

AH265

Product problems

  • Moisture or Humidity Problem
  • Moisture or Humidity Problem

Patient

Not reported

  • Increased Respiratory Rate
  • Increased Respiratory Rate

Narrative

Additional Manufacturer Narrative

AT THIS TIME, VYAIRE MEDICAL HAS NOT RECEIVED THE SAMPLE FOR INVESTIGATION. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

FELLOW M.D. REPORTED: TOO MUCH RAIN-OUT CAUSING VENT TO REGISTER HIGH RESPIRATORY RATE. ASKED TO HAVE CIRCUIT CHANGED OUT.

Additional Manufacturer Narrative

CODE 81 OTHER: THE DEVICE SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION BY OUR QUALITY TEAM. HOWEVER, THE QUALITY TEAM IDENTIFIED THAT A DEFECTIVE CRIMPING COULD CAUSE BAD ELECTRICAL CONNECTION OF HEATING WIRE TO THE POWER SUPPLY. THIS COULD GENERATE RAIN OUT IN THE CIRCUIT. IF SET UP TO CRIMPING EQUIPMENT IS NOT PROPERLY DONE, THE CRIMPING HEIGHT CAN BE OUT OF SPEC.