inforMED
MalfunctionOGL

INFANT HEATED WIRE CIRCUIT

Received Aug 8, 2018 · Event occurred Jun 24, 2018

Report 8030673-2018-00010 · MDR key 7761739

Device

Generic name

Oxygen Administration Kit

Manufacturer

Vyaire Medical

Catalog number

AH265

Lot number

0001119873

Product problems

  • Moisture or Humidity Problem
  • Moisture or Humidity Problem

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

OTHER-AT THIS TIME, THE SUSPECT COMPONENT IS NOT AVAILABLE FOR RETURN. DUE TO THE PART NOT BEING AVAILABLE TO BE RETURNED, NO FURTHER EVALUATION CAN BE COMPLETED AT THIS TIME.

Description of Event or Problem

THE RESPIRATORY THERAPIST REPORTED: EXCESSIVE RAIN-OUT.

Additional Manufacturer Narrative

THE DEVICE SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION BY OUR QUALITY TEAM. HOWEVER, THE QUALITY TEAM IDENTIFIED THAT A DEFECTIVE CRIMPING COULD CAUSE BAD ELECTRICAL CONNECTION OF THE HEATING WIRE TO THE POWER SUPPLY. THIS COULD GENERATE RAIN OUT IN THE CIRCUIT. IF SET UP TO CRIMPING EQUIPMENT IS NOT PROPERLY DONE, THE CRIMPING HEIGHT CAN BE OUT OF SPEC. THIS ISSUE WILL BE INTERNALLY INVESTIGATED WITHIN VYAIRE.