CLEAR-TRAC COMPLETE 5.5MM X 72MM
Received Aug 10, 2006 · Event occurred Mar 20, 2006
Report 1219602-2006-00115 · MDR key 773688
Device
Generic name
Search Of Records Shows
Manufacturer
Smith & Nephew, Inc., Endoscopy DivisionModel number
7209987Catalog number
7209987
Lot number
UNK
Product problems
- Break
- Component(s), broken
- Replace
- Tip breakage
Patient
38 YR
- Surgery, prolonged
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
A PIECE DID BREAK OFF INSIDE BODY OF PT AND THEY WERE ABLE TO REMOVE MOST OF IT. SURGEON INTRODUCED THE DISPOSABLE CANNULA, PUSHING THE CANNULA INTO THE SHOULDER JOINT. WHEN THE CANNULA WAS IN PLACE, HE REMOVED THE GUIDING ROD FROM THE POSTERIOR PORTAL AND RE-INSERTED THE TELESCOPE TO CARRY ON WITH THE PROCEDURE. IT WAS THEN DISCOVERED THAT, THE TIP PORTION OF THE DISPOSABLE WAS BROKEN. A DELAY OF THIRTY MINUTES RESULTED. BACK-UP DEVICES WERE AVAILABLE.
Additional Manufacturer Narrative
EVALUATION OF THE DEVICE SHOWED THE DISTAL END OF THE CANNULA IS BROKEN AND SMALL PIECES ARE MISSING. CONDITION OF THE DEVICE IS CONSISTENT WITH EXCESSIVE FORCE BEING APPLIED HOWEVER, A CONCLUSION COULD NOT BE MADE FOR THE REPORTED FAILURE.