inforMED
MalfunctionFBO

INNER TUBE WITH CERAMIC BEAK

Received Jun 22, 2018 · Event occurred May 31, 2018

Report 9610617-2018-00042 · MDR key 7631898

Device

Generic name

Resectoscope Sheath Inner Tube

Model number

27050XA

Catalog number

27050XA

Lot number

TU10

Product problems

  • Component Falling
  • Crack
  • Material Separation
  • Material Separation

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE SUBJECT DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION. THE DEVICE IS APPROXIMATELY 3 YEARS OLD; THE MOST LIKELY CAUSE OF THIS TYPE OF BREAKAGE IS MECHANICAL FORCE.

Description of Event or Problem

ALLEGEDLY, DURING AN INTRA-UTERINE FIBROID RESECTION PROCEDURE, THE TIP OF INNER TUBE CRACKED AND PIECES BROKE OFF INTO THE PATIENT. DOCTOR WAS ABLE TO RETRIEVE THE PIECES AND CONFIRMED THERE WAS NO PATIENT IMPACT.

Additional Manufacturer Narrative

SUPPLEMENTAL REPORT FILED TO CORRECT THE EVENT DATE FROM (B)(6) 2018 TO (B)(6) 2018.

Additional Manufacturer Narrative

WE EVALUATED THE RETURNED DEVICE AND FOUND THE CERAMIC BEAK IS BROKEN OFF FROM THE DISTAL END OF INNER TUBE AND THERE ARE WHAT APPEARS TO BE TOOL MARKS ON THE DISTAL END OF THE SHAFT. DAMAGE IS MOST LIKELY DUE TO THE CERAMIC BEAK MAKING CONTACT WITH A HARD SURFACE CAUSING AN INTERNAL CRACK AND RESULTING IN BREAKAGE; WE CANNOT CONFIRM.