inforMED
OtherMOS

MODEL #800239

Received Sep 11, 2006 · Event occurred Apr 19, 2006

Report 2183683-2006-00001 · MDR key 761355

Device

Generic name

7 Ch Bbc

Manufacturer

Invivo Corporation

Model number

800239

Catalog number

*

Lot number

*

Product problems

  • Unknown (for use when the device problem is not known)

Patient

UNKNOWN

  • Blister(s)
  • Bruise/Contusion
  • Burn(s)

Narrative

Description of Event or Problem

MRI COIL, 800239. "TWO PATIENTS RECEIVED MILD SMALL RF BURN TO RIGHT THUMB AND BRUISE UNDER LEFT THUMB WITH SAME COIL. BURN WAS SMALL BLISTER, SMALLER THAN THE DIAMETER OF A PEA." SITE STARTED PADDING PATIENTS HANDS ON LATER SCANS AND HAS STATED THEY DID APPROX. 90 SCANS WITHOUT INCIDENT AFTERWARDS.

Additional Manufacturer Narrative