Received Sep 8, 2006 · Event occurred Sep 6, 2006

Report 2954730-2006-00491 · MDR key 759522

Device

Generic name

Prothromin Time Test

Manufacturer

Hemosense, Inc.

Model number

*

Catalog number

*

Lot number

060218

Product problems

  • Incorrect Or Inadequate Test Results

Patient

*

  • Hospitalization required

Narrative

Description of Event or Problem

CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE 2006, INRATIO 2.1(THU) 1.5, LAB 8.0(FRI) 1.9 PATIENT WAS HOSPITALIZED.

Additional Manufacturer Narrative

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END USER AT TIME COMPLAINT WAS FILED: DATE 2006, INRATIO 2.10THU) 1.5, LAB 8.0 (FRI) 1.9, MEAN 5.05 1.7, CONFIDENCE LIMITS CANNOT BE DETERMINED 1.2-2.3. PER INTERNAL PROCEDURE, TR, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. FOR THE FIRST SET OF DATA, THE CONFIDENCE LIMITS CANNOT BE DETERMINED. THE TIME INTERVAL BETWEEN TWO INRATIO AND LAB TEST WAS >3HOURS. PER TR, THE COMPARISON CONSIDERED INVALID. FOR THE SECOND SET OF DATA, BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. PER TR, THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THIS IS AN ADVERSE EVENT CASE. PRODUCT WILL BE TESTED WHEN RETURNED.