MalfunctionGAI

VARADY PHLEBECTOMY EXTRACTOR, 6-3/4 SM HOOK

Received Jun 10, 2018 · Event occurred Apr 11, 2018

Report 9611112-2018-00003 · MDR key 7585423

Device

Generic name

Extractor

Manufacturer

Hebumedical Gmbh

Catalog number

10310

Lot number

100088-1711

Product problems

  • Fracture
  • Fracture

Patient

Not reported

  • No Consequences Or Impact To Patient
  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED WITHOUT THE BROKEN-OFF TIP. IT HAD BEEN MANUFACTURED IN NOVEMBER 2017. REVIEW OF MANUFACTURING RECORDS CONFIRMED DEVICE CONFORMANCE TO SPECS AT TIME OF RELEASE. HARDNESS OF RETURNED DEVICE WAS TESTED AND FOUND CONFORMING TO SPECIFICATIONS. THE DEVICE WAS SUBJECT TO MICROSCOPIC EXAMINATION. IT SHOWED NO EVIDENCE OF EXTENSIVE REUSE. SINCE THE DEVICE WAS RETURNED INCOMPLETE, I.E., WITHOUT THE BROKEN-OFF TIP, IT IS NOT POSSIBLE TO REACH A DEFINITE CONCLUSION WITH REGARD TO THE CAUSE OF THE FRACTURE. HOWEVER, THIS TYPE OF EXTRACTOR HAS A VERY DELICATE HOOK TIP AND THE TIP MAY FRACTURE IF EXCESSIVE PRESSURE IS APPLIED BY THE SURGEON.

Description of Event or Problem