inforMED
MalfunctionBWN

LITHOTOMY LEG HOLDER

Received Jun 8, 2018 · Event occurred May 11, 2018

Report 1043572-2018-00043 · MDR key 7584059

Device

Generic name

Lithotomy Leg Holder

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

Not reported

  • Fall
  • Fall

Narrative

Additional Manufacturer Narrative

THE LITHOTOMY LEG HOLDER IS EQUIPPED WITH A STIRRUP THAT SUPPORTS THE PATIENT'S LEG. THE STIRRUP IS SECURED WITH A SNAP HOOK CONNECTOR THAT CLIPS ONTO A RING. THE USER FACILITY PROVIDED PHOTOGRAPHS OF THE LITHOTOMY LEG HOLDER SUBJECT OF THE REPORTED EVENT. THE PHOTOGRAPHS WERE REVIEWED BY STERIS QUALITY AND PRODUCT ENGINEERING PERSONNEL AS PART OF THE INVESTIGATION AND REVEALED THAT THE STIRRUP'S CLIP DISENGAGED FROM THE RING DUE TO A GAP LARGE ENOUGH FOR THE SNAP HOOK CONNECTOR TO PASS THROUGH. BY DESIGN, THE SNAP HOOK ASSEMBLY IS ASSEMBLED IN A MANNER SUCH THAT THE SIZE OF THE GAP IN THE RING IS NOT SUFFICIENT TO ALLOW THE SNAP HOOK TO PASS THROUGH AS WAS SEEN IN THIS EVENT. A REVIEW OF ALL ON-HAND SNAP HOOK ASSEMBLIES WAS COMPLETED BY STERIS QUALITY PERSONNEL AND WERE FOUND TO BE ASSEMBLED PER SPECIFICATION. A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE EVENT AND REPLACED THE USER FACILITY'S SNAP HOOK ASSEMBLY. WHILE ONSITE, THE TECHNICIAN INSPECTED THE FACILITY'S OTHER LITHOTOMY LEG HOLDER ASSEMBLIES AND ENSURED NO GAP WAS PRESENT IN THE RING. THE TECHNICIAN TESTED THE LITHOTOMY LEG HOLDER ASSEMBLIES FOLLOWING HIS REPAIR, CONFIRMED THE EQUIPMENT WAS OPERATING TO SPECIFICATION AND RETURNED THE UNITS TO SERVICE. THIS EVENT IS ISOLATED. NO OTHER EVENTS OF THIS TYPE HAVE EVER BEEN REPORTED BEFORE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem

THE USER FACILITY REPORTED THAT DURING PATIENT PREPARATION PRIOR TO THE START OF THE PROCEDURE, THE PATIENT'S LEG BEGAN TO FALL FROM THE TABLE'S LITHOTOMY LEG HOLDER. USER FACILITY STAFF WERE ABLE TO CATCH THE PATIENT'S LEG; HOWEVER, THE WEIGHT OF THE LEG FALLING CAUSED THE DOCTOR'S ARM TO CONTACT THE SIDE OF THE SURGICAL TABLE RESULTING IN A SCRATCH. NO MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED FOR THE REPORTED SCRATCH.