SCALPEL 17.7*15.5*1.2*0.5 F/PIEZOELECTRI
Received Jun 6, 2018 · Event occurred Dec 12, 2017
Report 8030965-2018-54167 · MDR key 7576477
Device
Generic name
Handle, Scalpel
Manufacturer
Depuy Synthes Products LlcCatalog number
03.000.407S
Lot number
0102278
Product problems
- Overheating of Device
- Overheating of Device
Patient
42 YR
- Burn, Thermal
- Burn, Thermal
Narrative
Additional Manufacturer Narrative
UDI: (B)(6) IS UNAVAILABLE AS THE PRODUCT MADE PRIOR TO COMPLIANCE DATE; (B)(4). (B)(6). THE DEVICE MANUFACTURE DATE AND THE MANUFACTURING LOCATION ARE CURRENTLY UNAVAILABLE. AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
THIS IS REPORT 3 OF 3 FOR THE SAME EVENT: IT WAS REPORTED FROM (B)(6) THAT DURING RHINOPLASTY, IT WAS OBSERVED THAT THERE WAS DE-EPITHELIZATION OF THE SKIN ON THE LEFT SIDE OF THE NOSE WHILE USING THE SAW DEVICE TO TAKE DOWN THE DORSAL HUMP. ACCORDING TO THE REPORTER, THE SAW DEVICE WAS BEING USED WITH THE IRRIGATION TUBING DEVICE THAT APPEARED TO BE WORKING APPROPRIATELY. IT WAS FURTHER REPORTED THAT LONG SCALPEL DEVICES WERE ALSO USED AT THE TIME OF THE EVENT. IT WAS REPORTED THAT THE SURGEON WAS VERY SURPRISED THAT THERE WAS A SOFT TISSUE INJURY THAT RESULTED IN A FULL THICKNESS BURN ON THE SIDE OF THE NOSE GIVEN THAT THE SAW DEVICE WOULD ONLY WORK IN CONTACT WITH MINERALIZED BONE. THE SURGEON REPORTED THAT HE WAS USING THE BURR WITH THE TRANSVERSE SAW DEVICE FOR REDUCING THE DORSUM AND THOUGHT THAT THE END OF THE SAW DEVICE APPEARED TO HAVE CAUSED THE BURN TO THE SKIN. AS A RESULT, THE PATIENT ENDED UP WITH A BURN SCARRING AND WOULD REQUIRE A SECOND OPERATION. THE REPORTER STATED THAT THE PIEZO SAW DEVICE WAS USED AS PART OF A RHINOPLASTY TO REDUCE THE DORSAL HUMP AND TO CARRY OUT THE LATERAL OSTEOTOMIES. THERE WAS NO OTHER PROCEDURE THAN THE RHINOPLASTY. IT WAS FURTHER REPORTED THAT THE SURGERY WAS NOT PROLONGED BECAUSE OF THIS EVENT AS THE BURN WAS NOTED QUITE LATER ON IN THE PROCEDURE. THE SURGEON FURTHER REITERATED THAT IT WAS DIFFICULT AT THIS STAGE TO KNOW THE DEGREE OF RECOVERY THAT COULD BE EXPECTED AS IT WOULD CERTAINLY REQUIRE MONITORING OVER THE NEXT TWO YEARS AND A REVISION SURGERY. ACCORDING TO THE SURGEON, THE PATIENT¿S CURRENT CONDITION WAS THAT THEY HAD A BURNT SCAR ON THE LEFT SIDE OF THEIR NOSE, WHICH WAS DARK, THIN, ATROPHIC AND EXTREMELY NOTICEABLE. IN THAT REGARD, THERE WAS A SEVERE PERMANENT AND LASTING IMPACT TO THE PATIENT. THE SURGEON FURTHER STATED THAT, THEY WERE IN THE PROCESS OF MANAGING THE PATIENT¿S SCAR AND THE PATIENT WOULD HAVE TO UNDERGO A SCAR REVISION AT SOME POINT IN THE FUTURE. IT WAS FURTHER REPORTED THAT AT THE TIME OF THE SURGERY, THERE WAS NOTHING UNUSUAL ABOUT THE SAW FUNCTION, THE SAW PIECES OR THE EFFICACY OF THE SAW IN TREATING THE OSTEOTOMY OR THE DORSAL HUMP. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WAS PATIENT HARM THAT RESULTED IN A MEDICAL INTERVENTION AND PROLONGED HOSPITALIZATION. UPON FURTHER FOLLOW UP WITH THE REPORTER, IT WAS DISCOVERED THAT THE SURGEON PERFORMED A REVISION SURGERY ON THE PATIENT ON (B)(6) 2018. THE PATIENT¿S OUTCOME WAS NOT YET KNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Additional Manufacturer Narrative
DURING SUBSEQUENT FOLLOW-UP WITH THE CUSTOMER ADDITIONAL INFORMATION WAS OBTAINED. THE REPORTER STATED THAT THE PROCEDURE CHOSEN FOR THE RHINOPLASTY WAS AN OPEN APPROACH. IT WAS FURTHER REPORTED THAT THE PATIENT RECENTLY HAD A REVISION SURGERY WHICH WAS IN THE EARLY STAGES. IT WAS REPORTED THAT IT CURRENTLY LOOKED AS THOUGH THE DEPRESSED ELEMENT OF THE SCAR WAS MUCH IMPROVED; HOWEVER, THE SCAR WAS STILL AT THIS STAGE, YOUNG, A LITTLE RED AND INFLAMED. THE REPORTER INDICATED THAT THEY BELIEVED IT WOULD TAKE AT LEAST SIX MONTHS BEFORE THEY WERE ABLE TO COMMENT ON HOW SIGNIFICANT THE IMPROVEMENT HAD BEEN. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Additional Manufacturer Narrative
THE MEDICAL ADVISER FOLLOWED UP WITH THE SURGEON AND ADDITIONAL INFORMATION WAS OBTAINED. THE SURGEON REPORTED THAT THE PATIENT WAS A (B)(6) FEMALE WHOM UNDERWENT AN ELECTIVE RHINOPLASTY. THE SURGEON FELT THAT THE PIEZO SYSTEM WAS REVOLUTIONISING THAT TYPE OF SURGERY, AS HE HAD BEEN IMPRESSED AT THE ABILITY TO REDUCE THE DORSAL HUMP. THE SURGEON REPORTED THAT MANY SURGEONS WERE NOW SWITCHING TO THIS SYSTEM. IT WAS FURTHER REPORTED THAT THE LESION WAS A FULL THICKNESS BURN ON THE LEFT SIDE OF THE NOSE, LATERAL TO THE DORSAL HUMP. THAT WAS MANAGED AS FULL THICKNESS BURN AND LEFT TO ESCHAR. OVER THE FOLLOWING MONTHS THE ESCHAR RESULTED A PROMINENT SCALLOPING OF THE LEFT SIDE OF THE NOSE AND THE PATIENT UNDERWENT A SECOND PROCEDURE. THE SURGEON REPORTED NOT HAVING BEEN ON A FORMAL TRAINING CADAVERIC BASED COURSE FOR THE INSTRUMENT, BUT, HAD BEEN TO A NUMBER OF SYMPOSIA WHERE ¿KOL¿S ¿ HAD SPOKEN POSITIVELY ABOUT IT AND HE UNDERTOOK HIS FIRST CASE UNDER THE OBSERVATION OF THE DEPUY SYNTHES REPRESENTATIVE. HE REPORTED THAT THE PATIENT WAS HIS SECOND CASE. ACCORDING TO THE SURGEON, GETTING EACH PROCEDURAL STEP WAS NOT POSSIBLE. HE REPORTED HAVING TO MAKE A SLIGHTLY LARGER INCISION THAN FOR TRADITIONAL METHODS, BUT, WAS HAPPY TO DO SO DUE TO THE GOOD REDUCTION IN THE DORSAL HUMP IT PROVIDED. HE DISSECTED DOWN AND FLARED OPEN THE SOFT TISSUES. THEN HE USED A CONTINUOUS SWEEPING MOTION TO SHAVE DOWN THE DORSAL HUMP AND USED CONTINUOUS IRRIGATION WHILST TO DO SO. IT WAS FURTHER REPORTED THAT THE INJURY OCCURRED WHILE USING THE DEVICE TO SHAVE THE DORSAL HUMP. THE SURGEON FURTHER REPORTED THAT HE BELIEVED HE USED THE SMALLER SCALPEL TIP FOR THE REMOVAL OF THE DORSAL HUMP. THE SURGEON WHEN ON TO EXPLAIN WHICH STEP AND WITH WHICH TIP THE BURN HAPPENED. HE REPORTED THAT HE SHAVED THE DORSAL HUMP BY MOVING THE INSTRUMENT (407S) BACK AND FORTH UNTIL DESIRED REDUCTION WAS ACHIEVED AND WHILE HE WAS DOING THIS, HE NOTICED SMOKE COMING FROM THE SOFT TISSUE (LIKE DIATHERMY SMOKE) FROM THE SIDE OF THE DORSAL HUMP. IT WAS REPORTED THAT ITS THEN THAT HE NOTICED THE FULL THICKNESS OF THE BURN IN THE LEFT SIDE OF THE NASAL SOFT TISSUE. THE SURGEON REPORTED THAT THERE WAS NO OBVIOUS REDUCTION IN IRRIGATION FLOW OR CESSATION OF IRRIGATION. HE QUERIED WHETHER DESPITE THE SAW BEING IN DIRECT CONTACT WITH BONE, IF THE EDGE OF THE TIP WAS IN CONTACT WITH SOFT TISSUE COULD A BURN BE CAUSED BECAUSE OF; (A)MECHANICAL VIBRATION OF THE TIP AGAINST THE SKIN, (B) HEATING UP OF THE TIP WHICH MAY BE IN CONTACT WITH SOFT TISSUE AND, (C) IRRIGATION DOESN¿T RUN SUFFICIENTLY INTO THE PERIPHERY WHERE THE SCALPEL TIP MAY BE IN CONTACT WITH SOFT TISSUE. REGARDING THE AMOUNT OF LIQUID USED FOR EACH STEP, IT WAS REPORTED THAT THE SURGEON WAS UNABLE TO QUANTIFY AS THEY DID NOT MEASURE HOW MUCH FLUID WAS USED DURING THE SURGERY. ACCORDING TO THE REPORTER, THE RECORDS OF PROCEDURE WAS NOT ABLE TO BE VIEWED AS NOTES AT ANOTHER HOSPITAL; THEREFORE, THEY WERE UNABLE TO QUANTIFY THE FLOW RATE OF THE IRRIGATION (THAT WAS UNLIKELY TO BE RECORDED). THE SURGEON REPORTED THAT HE REPORTED NO OBVIOUS REDUCTION OR NOTICEABLE ISSUE WITH THE IRRIGATION WHICH HE REPORTED USING CONTINUOUSLY. THE SURGEON REPORTED THAT THE TIPS HAD BEEN USED ONCE AS THEY HAD THEM AS SINGLE USE INSTRUMENTS. IT WAS REPORTED THAT THE DEVICES WERE NOT AVAILABLE FOR INVESTIGATION BECAUSE THEY HAD BEEN DISPOSED OF AND NOT POSSIBLE TO RETRIEVE. THE SURGEON WAS PRAGMATIC THAT WITHOUT THE DEVICE / TIP IT WAS NOT POSSIBLE TO KNOW IF IT WAS USED IN ERROR, OR, IF THERE WAS AN ISSUE WITH THAT TIP. HE REPORTED THAT HE CONTINUED TO USE THE HAND DEVICE AND HAD NO ISSUES. HE HAD ALSO HAD NO FURTHER ISSUES WHILE USING THOSE TIPS; ALTHOUGH HE HAD NOW BECOME SELECTIVE WITH WHICH PATIENTS HE USES THE DEVICE WITH, DUE TO THIS EXPERIENCE. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Additional Manufacturer Narrative
THE DEVICE WAS NOT RETURNED FOR INVESTIGATION; THEREFORE, IT WAS NOT POSSIBLE TO ASSESS THE PROPER FUNCTIONALITY OF THE DEVICE, NEITHER TO CHECK THE PROPER MAINTENANCE OF THE PRODUCT. THE USED TIPS FOR THE PIEZO SYSTEM WERE DISPOSED AT THE CUSTOMER SITE. THE INVESTIGATION WAS PERFORMED BASED ON THE INFORMATION PROVIDED BY THE CUSTOMER AND IN COLLABORATION AND SUPPORT OF THE MANUFACTURER¿S PIEZO SYSTEM. THE ROOT CAUSE OF THE BURNED SOFT TISSUES COULD NOT BE FULLY DETERMINED BECAUSE THE TIP USED DURING THE SURGERY WAS NOT RECEIVED FOR INVESTIGATION. THEREFORE, THE INVESTIGATION WAS BASED UPON THE INFORMATION PROVIDED BY THE SURGEON. LABELING REVIEW: THE PIEZOELECTRIC SYSTEM INSTRUCTIONS FOR USE (IFU) INDICATES THAT ¿TO AVOID ANY RISK OF BURNS OR OVERHEATING, IRRIGATION MUST ALWAYS BE SWITCHED ON AND ACTIVE WHEN THE HANDPIECE IS IN USE. THE USER SHALL REGULARLY CONTROL THE LEVEL OF IRRIGATION TO ENSURE ITS COOLANT ACTION TO THE TIP. A LACK OF IRRIGATION PRESENTS A HAZARD TO THE PATIENT. DURING SURGERY IT SHOULD BE RESPECTED THAT THE SKIN IS NOT IN CONTACT WITH THE TIP AND PREVENTS THE IRRIGATION WHICH FINALLY HEATS UP THE TIP. ALTHOUGH THE DEVICE WAS NOT RETURNED FOR EVALUATION, THE AVAILABLE DATA RECEIVED FROM THE SURGEON¿S INTERVIEW SUPPORTED AND CONFIRMED THE REPORTED CONDITION. HOWEVER, AN ASSIGNABLE THE ROOT CAUSE COULD NOT BE FULLY DETERMINED AS NO FAILURES WERE IDENTIFIED SINCE THE PRODUCT WAS NOT RETURNED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.