inforMED
MalfunctionDSQ

HEARTWARE LVAD CONTROLLER

Received May 31, 2018 · Event occurred Apr 13, 2018

Report 7556773 · MDR key 7556773

Device

Generic name

Heartware Lvad Controller

Manufacturer

Minnetronix, Inc.

Catalog number

UNKNOWN

Lot number

UNKNOWN

Product problems

  • Loss of Power

Patient

27 YR

  • No Information
  • No Information

Narrative

Description of Event or Problem

PATIENT CALLED AND STATING THAT SHE HAD A "COMPLETE LOSS OF POWER" ON HER HEARTWARE LVAD CONTROLLER 2.0. SHE WAS CONNECTED TO TWO BATTERIES THAT HAD CHARGE. SHE WAS ABOUT TO HOOK TO HER AC UNIT. PATIENT DENIES ANY ISSUES WITH PUMP STOP. SHE WAS BROUGHT TO CLINIC AND LOG FILES SENT TO HEARTWARE/MEDTRONIC, WHICH CONFIRMED PUMP STOP. PATIENT GIVEN TWO NEW BATTERIES TO REPLACE THE ONES SHE WAS WEARING AT THE TIME.