VITEK® 2 ANC TEST KIT
Received May 17, 2018
Report 1950204-2018-00188 · MDR key 7522627
Device
Product problems
- Incorrect Or Inadequate Test Results
- Incorrect Or Inadequate Test Results
Patient
Not reported
- No Patient Involvement
- No Patient Involvement
Narrative
Description of Event or Problem
A CUSTOMER FROM ITALY REPORTED A MISIDENTIFICATION OF A CLOSTRIDIUM SPOROGENES (ATCC® 11437) INTERNAL QUALITY CONTROL STRAIN AS LOW DISCRIMINATION ORGANISM IN ASSOCIATION WITH THE VITEK® 2 ANC TEST KIT. THE CUSTOMER STATED THAT THE ANC CARD GAVE A LOW DISCRIMINATION ORGANISM RESULT FIVE TIMES. THE CUSTOMER HAD PERFORMED ONE QC WHICH WAS COMPLIANT. THE CUSTOMER USED CBA MEDIUM FOR 24 HOURS THEN SUBCULTURED AND INOCULATED THE STRAIN ON ANC CARDS. THE GRAM STAIN WAS GRAM POSITIVE BACILLI. THE CUSTOMER HAD CLEANED THE OPTICS AND THE VITEK SYSTEM DID NOT GIVE ANY ERRORS. THERE WAS NO PATIENT INVOLVEMENT AS THE EVENT PERTAINED TO A QUALITY CONTROL STRAIN. THE INSTRUCTIONS FOR USE (REF.043907-02), INDICATES THE ATCC 11437 C. SPOROGENES STRAIN IS NOT RECOMMENDED FOR ANC CARDS. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.
Additional Manufacturer Narrative
AN INTERNAL BIOMÉRIEUX INVESTIGATION WAS INITIATED DUE TO A LOW DISCRIMINATION RESULT WITH QUALITY CONTROL STRAIN CLOSTRIDIUM SPOROGENES ATCC 11437 ON VITEK® 2 V7.01 ANC CARD. THE CUSTOMER TESTED A STRAIN THAT IS NOT A BIOMÉRIEUX RECOMMENDED STRAIN FOR ANC QUALITY CONTROL. THE EXPECTED IDENTIFICATION TO CLOSTRIDIUM SPOROGENES WAS CONFIRMED ON VITEK® MS V3 (KB V3.0). ON VITEK 2 (V7.01) ANC CARDS, ONE (1) CARD OF CUSTOMER LOT (CL : 2440223403) AND ONE (1) CARD OF RANDOM LOT (RL : 2440320203) WERE TESTED FROM COS SUBCULTURE WITH ANAEROBIC ENVIRONMENT. THESE TESTS GAVE: -LOW DISCRIMINATION BETWEEN CL. BIFERMENTANS AND CL GROUP WITH THE CL. -UNIDENTIFIED ORGANISM WITH THE RL. CONCLUSION: THE CUSTOMER RESULT WAS REPRODUCED IN-HOUSE ON THE CL, WITH LOW REACTIVITY OF THE STRAIN: ATYPICAL STRAIN.