KRAYENBUEHL NERVE HOOK BALL-TIP 185MM
Received May 15, 2018 · Event occurred Mar 14, 2018
Report 9610612-2018-00209 · MDR key 7516286
Device
Generic name
Micro Instruments
Manufacturer
Aesculap AgModel number
FD397RCatalog number
FD397R
Lot number
52055649
Product problems
- Break
- Break
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
EXEMPTION NUMBER: E2014018. MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING.
Description of Event or Problem
(B)(6). THE CUSTOMER FOUND THAT THE WORKING END WAS BROKEN AND A FRAGMENT WAS MISSING AFTER AN OPERATION (B)(6) 2018. THE FRAGMENT COULD NOT FIND, THEREFORE, THERE IS THE POSSIBILITY OF REMAINING INSIDE PATIENT.
Additional Manufacturer Narrative
INVESTIGATION: THE INVESTIGATION WAS PERFORMED USING A KEYENCE VHX-5000 DIGITAL MICROSCOPE. THE ANALYSIS OF THE FRACTURE PATTERN ILLUSTRATED A FORCED FRACTURE DUE TO OVERLOAD. AN INDICATOR FOR AN OVERLOAD SITUATION IS THE BENT REMAINING PART OF THE WORKING END, MOST LIKELY CAUSED BY LEVERAGE. NO PORES, INCLUSIONS OR FOREIGN BODIES COULD BE FOUND ON THE POINT OF RUPTURE. BATCH HISTORY REVIEW: THE DEVICE QUALITY AND MANUFACTURING HISTORY RECORDS HAVE BEEN CHECKED FOR ALL AVAILABLE LOT NUMBERS AND BEEN FOUND TO BE ACCORDING TO OUR SPECIFICATION VALID AT THE TIME OF PRODUCTION. NO SIMILAR INCIDENTS HAVE BEEN FILED WITH PRODUCTS FROM THIS BATCH. CONCLUSION AND ROOT CAUSE: BASED ON THE INFORMATION AVAILABLE AS WELL AS A RESULT OF OUR INVESTIGATION, THE ROOT CAUSE OF FAILURE IS MOST PROBABLY RELATED TO AN INSUFFICIENT USAGE. RATIONAL: THIS KIND OF INSTRUMENT IS DESIGNED FOR DELICATE USE ONLY. BASED ON THE BENT WORKING END, IT IS ALMOST CERTAIN THAT A MECHANICAL OVERLOAD SITUATION LED TO THE BREAKAGE. NO CAPA IS NECESSARY.