WARMER DRAPE
Received May 7, 2018
Report 8043817-2018-00015 · MDR key 7492348
Device
Generic name
Fluid Warming And Slush Drapes
Manufacturer
Microtek Dominicana, S.a.Model number
ORS-300Lot number
D173059
Product problems
- Hole In Material
- Hole In Material
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
A SAMPLE WAS RECEIVED FOR INVESTIGATION. THE SAMPLE DRAPE WAS PLACED OVER A WARMER AND WATER WAS ADDED. THE DRAPE WAS LEFT ON THE WARMER FOR OVER 1 HOUR. NO LEAKS, TEARS, HOLES OR BURN MARKS WERE FOUND. THE DHR WAS REVIEWED FOR LOT D173059 AND IT WAS NOTICED THAT THIS LOT HAD 2736 PIECES THAT WERE MANUFACTURED ON 11/02/2017. NO DEFECTS WERE REPORTED DURING QUALITY INSPECTIONS. BASED ON THE DEVICE HISTORY RECORD AND SAMPLE REVIEW, THIS DOES NOT APPEAR TO BE THE RESULT OF A PERSONNEL, PROCESS, OR MATERIAL ISSUE AS NO LEAKS, TEARS, HOLES OR BURN MARKS WERE FOUND. SINCE THE REPORTED ISSUE WAS NOT FOUND REVIEW OF THE SAMPLE AND NO DEFECTS WERE REPORTED DURING QUALITY INSPECTION, NO ADDITIONAL ACTIONS WILL BE TAKEN AT THIS TIME.
Description of Event or Problem
COMPANY REP REPORTED "DRAPE IS TEARING DURING CASES. DID NOT SPECIFY ORIGINS OF TEARING BUT THAT THEY CONSIDERED DOUBLE DRAPING AS A SOLUTION. I DID LET THEM KNOW THIS IS NOT A GOOD IDEA AND COULD LEAD TO MORE ISSUES." NO PATIENT INJURY OR TREATMENT WAS REPORTED.