Received Jul 27, 2006 · Event occurred Jul 24, 2006

Report 1221934-2006-00155 · MDR key 746231

Device

Generic name

Shoft Tissue Fixation Device

Manufacturer

Depuy Mitek

Model number

210704

Catalog number

210704

Lot number

UNK

Product problems

  • Device, or device fragments remain in patient
  • Tip breakage

Patient

UNKNOWN

  • Unknown (for use when the patient's condition is not known)
  • Nonresorbable materials, unretrieved in body

Narrative

Description of Event or Problem

OUR AFFILIATE IS REPORTING THAT DURING A SHOULDER PROCEDURE ON OF THE 3 FIXATION DEVICES HAD THE TIP OF THE INSERTER MECHANISM BREAK OFF INTO THE ANCHOR AND REMAINS THEREIN THE ANCHOR WITHIN THE BODY. THE CASE WAS CONCLUDED SUCCESSFULLY WITHOUT FURTHER INCIDENT.

Additional Manufacturer Narrative

IT IS NOT KNOWN WHICH OF THE 3 DEVICE LOTS USED IN THE PROCEDURE IS THE ACTUAL DEVICE RELATIVE TO THE REPORTED INCIDENT, SO A BATCH REVIEW FOR ALL 3 LOTS WAS CONDUCTED TO DETERMINE IF THERE WERE ANY INTERNAL PROCESSING ISSUES WHICH WOULD HAVE CONTRIBUTED TO THE NATURE OF THE PRODUCT COMPLAINT. OUR RESULTS INDICATE THAT THESE 3 BATCHES OF PRODUCT WERE PROCESSED WITHOUT INCIDENT AND THEREFORE THERE IS NO INTERNALLY ASSIGNABLE CAUSE FOR THE REPORTED PROBLEM. FURTHER, A REVIEW INTO THE MITEK COMPLAINTS SYSTEM REVEALED NO OTHER COMPLAINTS OF ANY KIND FOR THESE 3 LOTS; LOT 0601154, LOT 0602170 AND LOT 062265 THAT WERE RELEASED TO DISTRIBUTION. ALTHOUGH THE COMPLAINT DEVICE HAS NOT YET BEEN RECEIVED FOR A FAILURE ANALYSIS, THIS TYPE OF FAILURE HAS HISTORICALLY BEEN ATTRIBUTED TO USER TECHNIQUE. FROM AN ENGINEERING PERSPECTIVE, DAMAGE TO THE INSERTER, TIP BREAKAGE, MAY HAVE BEEN CAUSED BY OF-ANGLE OR OFF AXIS INSERTION INTO THE BONE HOLE. OFF ANGLE/AXIS INSERTION IS THE MOST LIKELY CAUSE FOR THE INSERTER DISTAL TIP FAILURE, USUALLY THE RESULT OF BENDING AND/OR TWISTING THE ANCHOR DURING DEPLOYMENT DUE TO ANCHOR/BONE HOLE MISALIGNMENT. THE IFU STATES NOT TO TWIST AND/OR BEND THE INSERTER UPON INSERTION AS THE ANCHOR, SUTURE AND/OR INSERTER TIP MAY BE DAMAGED. WHEN THE DEVICE IS RECEIVED, IT WILL BE FORWARDED TO THE DEPUY MITEK QAE AND R&D GROUPS FOR A FAILURE ANALYSIS. IF THE RESULTS OF SAID INVESTIGATION DIFFER FROM THE ABOVE HYPOTHESIS A FOLLOW UP REPORT REFLECTING THE FAILURE ANALYSIS CONCLUSIONS WILL BE FILED.