inforMED
InjuryIME

THERAPEARL BREAST PACK

Received Apr 23, 2018 · Event occurred Feb 23, 2018

Report 1519375-2018-00002 · MDR key 7451510

Device

Generic name

Hot/cold Gel Compress

Catalog number

TP-LANSINOH

Lot number

02537-01

Product problems

  • Human Factors Issue
  • Human Factors Issue

Patient

Not reported

  • Burn(s)
  • Burn, Thermal
  • Burn(s)
  • Burn, Thermal

Narrative

Additional Manufacturer Narrative

INITIAL REPORT WAS RECEIVED 3/16/18. ON 3/19/18 THE HYGENIC TEAM DECIDED BASED ON THE INITIAL INFORMATION RECEIVED THAT IT WAS NOT A REPORTABLE EVENT. WE THEN RECEIVED FURTHER INFORMATION ON 3/29/18. THE TEAM MET AGAIN TO REVIEW THE ADDITIONAL INFORMATION AND DETERMINED IT NOW QUALIFIES AS A REPORTABLE EVENT. THE PRODUCT IS INTENDED TO BE HEATED, HOWEVER THE PATIENT HEATED THE PRODUCT INSIDE THE CLOTH COVER CAUSING THE BUTTON ON THE COVER TO BECOME HOT. THE PATIENT DOES NOT HAVE FULL FEELING IN HER BREAST AND DID NOT DETECT THE BURNING SENSATION RIGHT AWAY. THE PRODUCT WAS RETURNED AND INSPECTED AND WAS FOUND TO MEET PRODUCT SPECIFICATION.

Description of Event or Problem

WE HAVE RECEIVED VARYING ACCOUNTS OF THE INCIDENT FROM THE PATIENT'S HUSBAND WITH RESPECT TO THE HEATING AND USE OF THE THERAPEARL BREAST PACK. INITIALLY, THE HUSBAND STATED THAT HE MICROWAVED THE PACK FOR "LESS THAN ONE MINUTE". IN A SUBSEQUENT COMMUNICATION, THE HUSBAND STATED THAT IT WAS HIS WIFE WHO MICROWAVED THE PACK, AND THAT SHE DID SO FOR THE TIME RECOMMENDED ON THE PACKAGE. THE HUSBAND ALSO ADVISED THAT HIS WIFE "MAY HAVE MICROWAVED IT IN THE [CLOTH] SLEEVE." THE HUSBAND REPORTED THAT HE WAS ABLE TO SAFELY HANDLE THE PACK AND CARRY IT FROM THE MICROWAVE TO THE CHAIR WHERE HIS WIFE WAS SITTING. HIS WIFE THEN ALLEGEDLY PLACED THE BREAST PACK INSIDE HER SHIRT ON HER BREAST ON THE SIDE WHERE SHE WAS EXPERIENCING A CLOGGED DUCT. SHORTLY AFTER PLACING THE PACK ON HER BREAST, THE PATIENT REALIZED IT WAS BURNING HER AND QUICKLY REMOVED THE PACK. THE HUSBAND STATED HIS WIFE BELIEVED IT WAS THE PLASTIC BUTTON ON THE CLOTH SLEEVE THAT COVERS THE PACK THAT WAS THE HOT COMPONENT WHICH BURNT HER OUTER BREAST. THE PATIENT HAS PREVIOUSLY HAD BREAST ENLARGEMENT SURGERY AND DID NOT HAVE FULL FEELING IN ALL OF HER BREAST AT THE TIME OF THE INCIDENT DUE TO TIGHT AND THIN SKIN AND SCAR TISSUE, WHICH IS WHY SHE DID NOT REALIZE THE TEMPERATURE OF THE PACK AND THAT IT WAS AFFECTING HER SKIN. THE PATIENT WENT TO HER PRIMARY CARE DOCTOR REGARDING THE BURN. PER THE HUSBAND, THE PRIMARY CARE DOCTOR STATED THAT IT WAS A SIGNIFICANT BURN AND REFERRED THE PATIENT TO A DERMATOLOGIST FOR FURTHER INTERVENTION.