MalfunctionDWQ

TRIVEX SYSTEM RESECTOR HANDPIECE

Received Apr 18, 2018 · Event occurred Mar 15, 2018

Report 1220948-2018-00016 · MDR key 7441576

Device

Generic name

Handpiece

Catalog number

7210387

Product problems

  • Device Inoperable
  • Device Inoperable

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

WE HAVE RECEIVED AND EVALUATED THE COMPLAINT DEVICE. HOWEVER, WE CANNOT REPLICATE THE REPORTED FAILURE SINCE THE DEVICE WAS FOUND TO BE FUNCTIONING NORMALLY. WHEN THE RUN AND WINDOW LOCK BUTTONS WERE PRESSED, THEY FUNCTIONED AS INTENDED. THE ENDBELL CONNECTOR OF THE CABLE WAS CLEAN WITHOUT ANY SIGNS OF WEAR. WHEN THE UNIT WAS DISSEMBLED, THERE WERE NO SIGNS OF DAMAGE TO THE INNER ASSEMBLY. AT THIS TIME, WE ARE INCONCLUSIVE ABOUT THE ROOT CAUSE OF THE DEFECT SINCE THE DEVICE FUNCTIONED NORMALLY DURING OUR EVALUATION. THERE WAS NO INJURY TO THE PATIENT AS THE RESULT OF THIS INCIDENT. THE PROCEDURE WAS COMPLETED USING A DIFFERENT HANDPIECE.

Description of Event or Problem

THE HANDPIECE SHUTS OFF DURING THE PROCEDURE.