ABL700

Received Jul 17, 2006 · Event occurred Jun 16, 2006

Report 3002807968-2006-00036 · MDR key 743489

Device

Generic name

Blood Gas Analyzer

Model number

ABL735

Catalog number

393-725

Lot number

*

Product problems

  • Device Alarm System
  • Low Test Results

Patient

*

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

A TOO LOW NA-VALUE WAS MEASURED ON AN ABL735 BLOOD GAS ANALYZER. NO ALARM WAS ISSUED. THE LOW NA VALUE MAY HAVE BEEN CAUSED BY A DEFECTIVE REFERENCE ELECTRODE. ALLEGEDLY 4 OR 5 PATIENTS WERE INAPPROPRIATELY TREATED FOR TOO LOW NA VALUES IN THE BLOODSTREAM. NO SERIOUS INJURY OR DEATH HAS BEEN REPORTED AS A CONSEQUENCE OF THESE TREATMENTS.

Additional Manufacturer Narrative

THE HOSPITAL ALLEGES THE REFERENCE MEMBRANE IS DRIFTING AFTER 4 OR 5 DAYS. ADDITIONAL TESTS ARE UNDERTAKEN.