inforMED
MalfunctionGET

ENDO DISSECT

Received Apr 13, 2018 · Event occurred Mar 20, 2018

Report 2647580-2018-01823 · MDR key 7429329

Device

Generic name

Motor, Surgical Instrument, Pneumatic Powered

Model number

176645

Catalog number

176645

Lot number

P7H1974X

Product problems

  • Failure to Align
  • Failure to Align

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Additional Manufacturer Narrative

POST MARKET VIGILANCE (PMV) LED AN EVALUATION OF SIX DEVICES. ENGINEERING VERIFIED THAT JAW ALIGNMENT OF DEVICES WAS WITHIN SPECIFICATIONS. ENGINEERING INSPECTED JAWS UNDER A MICROSCOPE AND FOUND NO PHYSICAL DAMAGE TO ANY OF THE JAWS. RECORDS FROM EACH MANUFACTURING LOT ARE THOROUGHLY REVIEWED TO ENSURE THAT PRODUCTS ARE RELEASED MEETING ALL QUALITY RELEASE SPECIFICATIONS AT THE TIME OF MANUFACTURE. ANALYSIS CONCLUDED THERE WERE NO ASSEMBLY COMPONENT RELATED FAILURES. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem

ACCORDING TO THE REPORTER, PRE OPERATIVELY, THE TIP OF THE DISSECTOR WAS NOT ALIGNING CORRECTLY. PRODUCT NOT UP TO STANDARD. THERE WAS NO PATIENT INVOLVEMENT.