SVC REPL 660HD-E IMAGE MANAGEMENT SYSTEM
Received Apr 13, 2018 · Event occurred Feb 21, 2018
Report 1643264-2018-00285 · MDR key 7428851
Device
Generic name
Device, Digital Image Storage, Radiological
Manufacturer
Smith & Nephew, Inc.Model number
72204668SCatalog number
72204668S
Product problems
- No Display/Image
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
IT WAS REPORTED THAT THE DEVICE HAD THE SCREEN BLACK. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO DELAY OR PATIENT INJURIES.
Additional Manufacturer Narrative
THE DEVICE WAS RECEIVED FOR EVALUATION. THERE WAS NO RELATIONSHIP FOUND BETWEEN THE RETURNED DEVICE AND THE REPORTED INCIDENT. A VISUAL INSPECTION WAS PERFORMED ON THE PRODUCT AND NO ISSUE WAS OBSERVED. COMPLAINT OF BLANK DISPLAY COULD NOT BE REPRODUCED. UNIT PASSED FUNCTIONAL TESTING DURING 6 HOUR BURN-IN INCLUDING CAPTURE AND PRINT FUNCTION. UNIT DID NOT LOSE DISPLAY OR TOUCHSCREEN FUNCTION DURING BURN-IN. ALL VIDEO INPUTS/OUTPUTS NORMAL AND UNIT'S SOFTWARE PERFORMED AS EXPECTED. UNIT WAS TESTED IN BOTH PAL AND NTSC FORMATS AND BOTH ANALOG AND DIGITAL VIDEO OUTPUTS. THE COMPLAINT WAS NOT CONFIRMED AND THE ROOT CAUSE COULD NOT BE DETERMINED SINCE THE REPORTED MALFUNCTION COULD NOT BE DUPLICATED DURING THE PRODUCT EVALUATION PROCESS. A REVIEW OF THE DEVICE HISTORY RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATION OR WOULD NOT BE ABLE TO PERFORM AS INTENDED.